Mass Cytometry: Technique for Real Time Single Cell Multitarget Immunoassay Based on Inductively Coupled Plasma Time-of-Flight Mass Spectrometry

2009· article· en· 1,331 citations· W2038254572 on OpenAlex· 10.1021/ac901049w

Why is this work in the frame?

A frame that forgets how it found something cannot be audited. These are the routes that admitted this work.

Canadian affiliationAn author listed a Canadian institution. This is the only route the usual frame has.
Canadian funderA Canadian agency funded it. The work may carry no Canadian affiliation at all.

Machine scores (provisional)

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Opus teacher head0.008
GPT teacher head0.253
Teacher spread
0.245 · how far apart the two teachers sit on this one work
Validation status
score_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Abstract

A novel instrument for real time analysis of individual biological cells or other microparticles is described. The instrument is based on inductively coupled plasma time-of-flight mass spectrometry and comprises a three-aperture plasma-vacuum interface, a dc quadrupole turning optics for decoupling ions from neutral components, an rf quadrupole ion guide discriminating against low-mass dominant plasma ions, a point-to-parallel focusing dc quadrupole doublet, an orthogonal acceleration reflectron analyzer, a discrete dynode fast ion detector, and an 8-bit 1 GHz digitizer. A high spectrum generation frequency of 76.8 kHz provides capability for collecting multiple spectra from each particle-induced transient ion cloud, typically of 200-300 micros duration. It is shown that the transients can be resolved and characterized individually at a peak frequency of 1100 particles per second. Design considerations and optimization data are presented. The figures of merit of the instrument are measured under standard inductively coupled plasma (ICP) operating conditions (<3% cerium oxide ratio). At mass resolution (full width at half-maximum) M/DeltaM > 900 for m/z = 159, the sensitivity with a standard sample introduction system of >1.4 x 10(8) ion counts per second per mg L(-1) of Tb and an abundance sensitivity of (6 x 10(-4))-(1.4 x 10(-3)) (trailing and leading masses, respectively) are shown. The mass range (m/z = 125-215) and abundance sensitivity are sufficient for elemental immunoassay with up to 60 distinct available elemental tags. When <15 elemental tags are used, a higher sensitivity mode at lower resolution (M/DeltaM > 500) can be used, which provides >2.4 x 10(8) cps per mg L(-1) of Tb, at (1.5 x 10(-3))-(5.0 x 10(-3)) abundance sensitivity. The real-time simultaneous detection of multiple isotopes from individual 1.8 microm polystyrene beads labeled with lanthanides is shown. A real time single cell 20 antigen expression assay of model cell lines and leukemia patient samples immuno-labeled with lanthanide-tagged antibodies is presented.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

The record

Venue
Analytical Chemistry
Topic
Advanced biosensing and bioanalysis techniques
Field
Biochemistry, Genetics and Molecular Biology
Canadian institutions
University of TorontoUniversity Health Network
Funders
Ruhr-Universität BochumOntario Genomics InstituteGenome Canada
Keywords
ChemistryMass spectrometryAnalytical Chemistry (journal)Inductively coupled plasma mass spectrometryDynodeReflectronInductively coupled plasmaIonQuadrupoleTime-of-flight mass spectrometryPlasmaAtomic physicsIonizationPhysicsAnode