Monte Carlo modelling of diode detectors for small field MV photon dosimetry: detector model simplification and the sensitivity of correction factors to source parameterization
Notice bibliographique
Résumé
The goal of this work was to examine the use of simplified diode detector models within a recently proposed Monte Carlo (MC) based small field dosimetry formalism and to investigate the influence of electron source parameterization has on MC calculated correction factors. BEAMnrc was used to model Varian 6 MV jaw-collimated square field sizes down to 0.5 cm. The IBA stereotactic field diode (SFD), PTW T60016 (shielded) and PTW T60017 (un-shielded) diodes were modelled in DOSRZnrc and isocentric output ratios (OR(fclin)(detMC)) calculated at depths of d = 1.5, 5.0 and 10.0 cm. Simplified detector models were then tested by evaluating the percent difference in (OR(fclin)(detMC)) between the simplified and complete detector models. The influence of active volume dimension on simulated output ratio and response factor was also investigated. The sensitivity of each MC calculated replacement correction factor (k(fclin,fmsr)(Qclin,Qmsr)), as a function of electron FWHM between 0.100 and 0.150 cm and energy between 5.5 and 6.5 MeV, was investigated for the same set of small field sizes using the simplified detector models. The SFD diode can be approximated simply as a silicon chip in water, the T60016 shielded diode can be modelled as a chip in water plus the entire shielding geometry and the T60017 unshielded diode as a chip in water plus the filter plate located upstream. The detector-specific (k(fclin,fmsr)(Qclin,Qmsr)), required to correct measured output ratios using the SFD, T60016 and T60017 diode detectors are insensitive to incident electron energy between 5.5 and 6.5 MeV and spot size variation between FWHM = 0.100 and 0.150 cm. Three general conclusions come out of this work: (1) detector models can be simplified to produce OR(fclin)(detMC) to within 1.0% of those calculated using the complete geometry, where typically not only the silicon chip, but also any high density components close to the chip, such as scattering plates or shielding material is necessary to be included in the model, (2) diode detectors of smaller active radius require less of a correction and (3) (k(fclin,fmsr)(Qclin,Qmsr)) is insensitive to the incident the electron energy and spot size variations investigated. Therefore, simplified detector models can be used with acceptable accuracy within the recently proposed small field dosimetry formalism.
Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.
Comment cette classification a été obtenuedéplier
Prédiction distillée sur la base complète
Imitation des enseignantsNi prévalence calibrée, ni vérité terrain. Validation humaine à venir. Apprise à partir de 10 348 étiquettes directes de Codex et de 10 348 étiquettes directes de Gemma. Le mode candidate est l'union des têtes enseignantes seuillées; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont ni des étiquettes humaines ni des étiquettes directes de modèles de pointe.
Scores Codex et Gemma par catégorie
| Catégorie | Codex | Gemma |
|---|---|---|
| Métarecherche | 0,000 | 0,000 |
| Méta-épidémiologie (sens strict) | 0,000 | 0,000 |
| Méta-épidémiologie (sens large) | 0,000 | 0,000 |
| Bibliométrie | 0,000 | 0,000 |
| Études des sciences et des technologies | 0,000 | 0,000 |
| Communication savante | 0,000 | 0,000 |
| Science ouverte | 0,000 | 0,000 |
| Intégrité de la recherche | 0,000 | 0,000 |
| Charge utile insuffisante (le modèle a refusé de juger) | 0,000 | 0,000 |
Scores machine (provisoires)
Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.
Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.
score_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découleClassification
machine, non validéePrédiction automatique; un appel candidat d’une seule tête enseignante, pas un consensus.
Le détail, modèle par modèle et score par score, se trouve en fin de page sous « Comment cette classification a été obtenue ».