Diffractive Optics Technology: A Novel Detection Technology for Immunoassays

2006· article· en· W2125264414 sur OpenAlexaff
Vitali Borisenko, Wei Hu, Pui Lam Tam, Irene Chen, Jean-François Houle, Walter A. Ausserer

Notice bibliographique

RevueClinical Chemistry · 2006
Typearticle
Langueen
DomaineChemical Engineering
ThématiqueAnalytical Chemistry and Sensors
Établissements canadiensAxela (Canada)
Organismes subventionnairesNational Cancer Institute
Mots-clésOpticsComputer sciencePhysics

Résumé

récupéré en direct d'OpenAlex

There is an increasing need for high-sensitivity immunoassays that can be used in point-of-care patient testing of complex media. For example, analytes such as the natriuretic peptides and recently discovered sepsis markers are found in blood in very low picomolar concentrations (1)(2). Although advances have been made in the use of fluorescent, chemiluminescent, and other labels to measure markers at lower detection limits, background interference from biological samples and detection instrumentation remains problematic. Optical biosensors offer the promise of label-free real-time measurements, but their application to quantification of analytes in complex media is impaired by higher detection limits and is susceptible to changes in refractive indices or nonspecific surface binding. Moreover, the costliness of these devices has largely prohibited bedside implementation. In this report, we detail the use of a novel diffractive optics technology (dot™) that takes advantage of the inherent properties of diffractive optics to deliver a cost-effective, portable, robust, optical biosensor that detects analytes at picomolar concentrations in complex media. In the dotLab™ System, coherent light striking a nonrandom pattern of capture molecules on the dotLab Sensor creates constructive and destructive interferences that produce a well-defined diffraction image. As molecules bind to the capture molecules, the height of the diffraction pattern is increased, which in turn increases the diffraction efficiency and the diffractive order intensity. A photodiode monitors the intensity of the diffractive order, which is correlated to analyte concentrations. Because diffraction is inherently self-referencing, the transduction of binding events is dependent on the initial pattern, and an increase in diffractive order intensity will occur only if molecules bind exclusively to the patterned capture reagents. Therefore, nonspecific binding to both the patterned and nonpatterned regions will not affect the signal, a characteristic that offers an important advantage over other optical biosensor systems in which any surface-binding event will cause an increase in signal. Previous diffraction-based immunosensors have used silicon wafer chips that were in contact with analyte-containing solutions and required washing and drying before analysis by a simple reader (3). The dotLab system uses a plastic consumable, the dotLab Sensor, with an integrated prism situated below the flow channel so that the light source interrogates the diffraction grating without passing through the bulk solution. Previous studies suggest that this total internal reflection scheme allows for 95% of the laser intensity to be measured, whereas in a nontotal internal reflection set-up only 5% is measured (4). Moreover, with this configuration, our system monitors biomolecule binding in real time and in complex media. We tested the beta prototype of the dotLab instrument, which incorporates the core dot technology in an integrated package intended for in-house and controlled external development work. The instrument includes precision fluidic control of reagents, buffers, and samples—a proprietary integrated optical assembly designed to function with the dotLab Sensor; and software for acquisition, control, and user-interface. A movable stage allows monitoring and potential patterning with different capture reagents for 8 discrete diffraction spots. In addition, each individual spot can be cross-patterned to perform intraspot multiplexing and assays, as we have previously demonstrated (5)(6). We used a single-spot and a streptavidin-patterned sensor with reagents for the detection of N-terminal probrain natriuretic peptide (NT-proBNP) in various matrices at reference concentrations, as summarized in Table 11 . For direct detection of recombinant NT-proBNP, we used a biotinylated monoclonal antibody directed against NT-proBNP immobilized on a streptavidin-patterned dotLab Sensor (Fig. 1A1 ). The binding event can be observed in real time in the trace. The binding of recombinant NT-proBNP is also detected, but at a lower intensity. The recombinant protein has a low relative molecular mass (Mr) of ∼8000, which generates very little signal on its own. At lower concentrations, we used a different format to enhance the diffraction signal. For example, to detect NT-proBNP at nanomolar concentrations, we used a combination of capture and detector antibodies (Fig. 1B1 ), with which the binding of the biotinylated mouse capture antibody was readily detected, whereas the introduction of the recombinant protein did not produce a detectable signal increase. We carried out rinse steps with pulses of phosphate buffered saline (PBS) (137 mmol/L NaCl; 2.7 mmol/L KCl; 10 mmol/L Na2HPO4/KH2PO4, pH 7.4; OmniPur) and PBS-Tween (10 mmol/L phosphate buffer, pH 7.4; 140 mmol/L NaCl; 3 mmol/L KCl; 0.025% (w/v) Tween-20; Calbiochem). When we introduced a polyclonal goat antibody directed against NT-proBNP, the diffraction signal increased. For internal calibration purposes and to minimize any impact of intersensor variability, we calculated the ratio of the signal increase for the detector-binding event to that for the capture-antibody binding event. With this approach we generated a calibration curve for recombinant NT-proBNP spanning 10 to 1000 μg/L, or ∼1.25 to 125 nmol/L in a PBS matrix (Fig. 1C1 ). For detection at picomolar concentrations, we added a 3rd antibody, donkey antigoat horseradish peroxidase conjugate, to the assay format. To investigate the compatibility of this system with a typical clinical matrix, we ran these assays in a human plasma matrix stripped of endogenous NT-proBNP. For these experiments, we incubated the recombinant NT-proBNP with all 3 antibodies (see Table 11 ) in a single mixture before loading onto the sensor surface. The 90-min preincubation was necessary because of the relatively low on-rate of 1 antibody in the mixture. The binding event detected after loading the mixture reflects the binding of free and complexed biotinylated monoclonal antibody. After being washed briefly in both PBS-Tween and PBS, the complexes were detected by incubation with a precipitating form of 3,3′,5,5′ tetramethylbenzidine (TMB) [1-component TMB membrane peroxidase substrate (Kirkegaard & Perry Laboratories)]. The specific and localized precipitation of TMB on the diffraction pattern causes a signal increase. We again performed internal calibration by use of the ratio of the signal generated by the TMB precipitation to the signal generated by the binding events (Fig. 1D1 ); alternatively, a ratio of binding rates can be used, obviating the need to observe reactions until saturation. We repeated this experiment over a range of calibrator concentrations and generated the calibration curve shown in Fig. 1E1 . Each point is the mean of a duplicate or triplicate determination. The curve spans 31.25 to 2500 ng/L (∼4–300 pmol/L). The lowest value reported is 2 SDs above the mean determination for the zero analyte control. Several clinically relevant decision points are highlighted in Fig. 1E1 . In conclusion, we demonstrated that the dotLab system can be used to detect a clinically relevant analyte over a wide range of concentrations in complex media. The observation of real-time binding can reduce the time required to obtain results, because quantification based on rate measurement is inherently more rapid than end-point determinations. For the assay developer, real-time binding observations help identify unwanted reagent interactions that may be sources of background interference or noise. The detection and capture molecules can be modified to quantify other biomolecules such as DNA; we have detected the hybridization of cDNA strands (not shown). Thus, our method lends itself to the detection of many types of analytes. The simple, robust optics technology at the core of the current compact bench-top platform can be readily deployed, as in the current stand-alone system, or incorporated into various other platforms in the central laboratory and for bedside patient testing. Experimental components and assay format. N/A, not applicable Experimental components and assay format. N/A, not applicable Detection and quantification of NT-proBNP over a range of concentrations. (A), real-time trace from the dotLab system, truncated because of space limitations. The dashed box is an enhanced view of the NT-proBNP binding to the immobilized capture antibody. Unprocessed voltage signal from the photodiode reflects the intensity of the diffractive order measured in diffraction units (DU). Arrows indicate introduction of capture antibody (CA) and NT-proBNP calibrator (NT). (B), a real–time trace from the 2 component assays. Bars indicate the approximate location where delta values were derived and used to establish the ratiometric value for each concentration of calibrator. Arrows indicate introduction of CA, NT, detector antibody (DT), and rinse step with PBS or PBS-Tween (R). (C), relationship of delta ratio to NT-proBNP calibrator concentrations. (D), real-time trace from the multicomponent assay. Arrows indicate introduction of mixture (CK) (see Table 11 ). TMB and R. The large DU increase during the R phase indicates a change in bulk refractive index of the solution from plasma to the PBS washing matrix that did not affect our ability to monitor the subsequent TMB precipitation reaction. Bars indicate the approximate location of the delta values used in the ratiometric determination. (E), relationship of delta ratio to NT-proBNP calibrator concentrations in plasma. Clinical decision points: ①, the clinical cutoff for congestive heart failure (CHF) (1), ② and ③, age-related cut-offs for acute CHF (7), ④, recently suggested cutoff for short-term adverse outcome (8).

Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.

Comment cette classification a été obtenuedéplier

Prédiction distillée sur la base complète

Imitation des enseignants

Ni prévalence calibrée, ni vérité terrain. Validation humaine à venir. Apprise à partir de 10 348 étiquettes directes de Codex et de 10 348 étiquettes directes de Gemma. Le mode candidate est l'union des têtes enseignantes seuillées; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont ni des étiquettes humaines ni des étiquettes directes de modèles de pointe.

score de la tête « metaresearch » (Codex)0,000
score de la tête « metaresearch » (Gemma)0,002
Version: codex-gemma-dda1882f352aStatut de validation: machine_predicted_unvalidated
Catégories candidatesMéta-épidémiologie (sens strict), Intégrité de la recherche
Catégories consensuellesaucune
DomaineSignal candidat: aucune · Signal consensuel: aucune
Devis d'étudeSignal candidat: Expérimental (laboratoire) · Signal consensuel: Expérimental (laboratoire)
GenreSignal candidat: Empirique · Signal consensuel: Empirique
Score de désaccord entre enseignants0,100
Score d'incertitude au seuil1,000

Scores Codex et Gemma par catégorie

CatégorieCodexGemma
Métarecherche0,0000,002
Méta-épidémiologie (sens strict)0,0000,000
Méta-épidémiologie (sens large)0,0000,000
Bibliométrie0,0000,000
Études des sciences et des technologies0,0000,000
Communication savante0,0000,000
Science ouverte0,0000,000
Intégrité de la recherche0,0010,001
Charge utile insuffisante (le modèle a refusé de juger)0,0000,000

Scores machine (provisoires)

Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.

Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.

Tête enseignante Opus0,019
Tête enseignante GPT0,297
Écart entre enseignants0,278 · la distance entre les deux têtes enseignantes sur ce seul travail
Statut de validationscore_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découle

Classification

machine, non validée

Prédiction automatique; un appel candidat d’une seule tête enseignante, pas un consensus.

Devis d'étudeExpérimental (laboratoire)
Domainenon disponible
GenreEmpirique

Le détail, modèle par modèle et score par score, se trouve en fin de page sous « Comment cette classification a été obtenue ».

En bref

Citations20
Publié2006
Routes d'admission1
Résumé présentoui

Explorer davantage

Même revueClinical ChemistryMême sujetAnalytical Chemistry and SensorsTravaux en français237 207