Simulated-use validation of a sponge ATP method for determining the adequacy of manual cleaning of endoscope channels
Notice bibliographique
Résumé
BACKGROUND: The objective of this study was to validate the relative light unit (RLU) cut-off of adequate cleaning of flexible colonoscopes for an ATP (adenosine tri-phosphate) test kit that used a sponge channel collection method. METHODS: This was a simulated-use study. The instrument channel segment of a flexible colonoscope was soiled with ATS (artificial test soil) containing approximately 8 Log10 Enterococcus faecalis and Pseudomonas aeruginosa/mL. Full cleaning, partial cleaning and no cleaning were evaluated for ATP, protein and bacterial residuals. Channel samples were collected using a sponge device to assess residual RLUs. Parallel colonoscopes inoculated and cleaned in the same manner were sampled using the flush method to quantitatively assess protein and bacterial residuals. The protein and viable count benchmarks for adequate cleaning were <6.4 ug/cm(2) and <4 Log10 cfu/cm(2). RESULTS: The negative controls for the instrument channel, over the course of the study remained low with on average 14 RLUs, 0.04 ug/cm(2) protein and 0.025 Log10 cfu/cm(2). Partial cleaning resulted in an average of 6601 RLUs, 3.99 ug/cm(2), 5.25 Log10 cfu/cm(2) E. faecalis and 4.48 Log10 cfu/cm(2) P. aeruginosa. After full cleaning, the average RLU was 29 (range 7-71 RLUs) and the average protein, E. faecalis and P. aeruginosa residuals were 0.23 ug/cm(2), 0.79 and 1.61 Log10 cfu/cm(2), respectively. CONCLUSIONS: The validated cut-off for acceptable manual cleaning was set at ≤100 RLUs for the sponge collected channel ATP test kit.
Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.
Comment cette classification a été obtenuedéplier
Prédiction distillée sur la base complète
Imitation des enseignantsNi prévalence calibrée, ni vérité terrain. Validation humaine à venir. Apprise à partir de 10 348 étiquettes directes de Codex et de 10 348 étiquettes directes de Gemma. Le mode candidate est l'union des têtes enseignantes seuillées; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont ni des étiquettes humaines ni des étiquettes directes de modèles de pointe.
Scores Codex et Gemma par catégorie
| Catégorie | Codex | Gemma |
|---|---|---|
| Métarecherche | 0,002 | 0,010 |
| Méta-épidémiologie (sens strict) | 0,000 | 0,000 |
| Méta-épidémiologie (sens large) | 0,000 | 0,000 |
| Bibliométrie | 0,000 | 0,000 |
| Études des sciences et des technologies | 0,000 | 0,000 |
| Communication savante | 0,000 | 0,000 |
| Science ouverte | 0,000 | 0,000 |
| Intégrité de la recherche | 0,000 | 0,000 |
| Charge utile insuffisante (le modèle a refusé de juger) | 0,000 | 0,000 |
Scores machine (provisoires)
Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.
Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.
score_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découleClassification
machine, non validéePrédiction automatique; un appel candidat d’une seule tête enseignante, pas un consensus.
Le détail, modèle par modèle et score par score, se trouve en fin de page sous « Comment cette classification a été obtenue ».