Evaluating Surgeon Estimation of Cup Position in Total Hip Arthroplasty: A Cadaver Study
Notice bibliographique
Résumé
Abstract Inaccurate placement of components during total hip arthroplasty (THA) can lead to significant postoperative complications including revision surgery. Traditionally, surgeons grossly estimate component positioning intraoperatively using anatomical landmarks; however, evidence indicates that this surgeon assessment may not be reliable. The purpose of this study was to determine the accuracy of surgeon estimates of component position as compared with imaging (radiographs and computed tomography [CT] scan) and a new surgical navigation system. Three board-certified orthopaedic surgeons each performed four THA procedures on six cadavers (12 hips). Radiographs and CT scans were obtained postoperatively. The “gold standard” measurements of implanted cup anteversion and inclination were derived from three-dimensional renderings created from postoperative CTs. A reference value for cup position was created by aligning the anterior pelvic plane in each rendering coplanar with the CT table. Following each procedure, surgeons provided their estimate of acetabular cup component orientation. Surgeon estimates were compared with data gathered from postoperative radiographs, CT scans, and the navigation device. Surgeon estimates of anteversion and inclination were within 10 degrees of reference values in 64% (7/11) and 82% (9/11) of cases, respectively. Surgeon estimates of anteversion differed from reference values by a mean of 7.6 ± 5 degrees, whereas inclination differed from reference values by a mean of 6.1 ± 5.1 degrees (all means absolute). Radiographic measurements differed from reference values by 7.8 ± 4.3 degrees (p > 0.05) and 2.7 ± 2.3 degrees (p = 0.06) for anteversion and inclination, respectively, whereas CT values differed by 2.5 ± 1.6 degrees (p = 0.004) and 2.3 ± 2.1 degrees (p = 0.04). The navigation system differed from reference values by 4 ± 4 degrees (p = 0.08) and 4.2 ± 3.2 degrees (p = 0.31). Surgeons underestimated anteversion and inclination by 7.7 ± 4.8 degrees and 6.9 ± 4.8 degrees, respectively. Surgeon underestimation was observed in 8/11 (73%) cases, with anteversion underestimated by > 5 degrees in 5/8 (62%) cases and inclination underestimated by > 5 degrees in 4/8 (50%) cases. Our findings suggest that surgeons tend to underestimate both anteversion and inclination and that the accuracy of their estimates is similar to that of radiographs. CT scans and the navigation system were able to provide more accurate measurements of cup position.
Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.
Comment cette classification a été obtenuedéplier
Prédiction distillée sur la base complète
Imitation des enseignantsNi prévalence calibrée, ni vérité terrain. Validation humaine à venir. Apprise à partir de 10 348 étiquettes directes de Codex et de 10 348 étiquettes directes de Gemma. Le mode candidate est l'union des têtes enseignantes seuillées; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont ni des étiquettes humaines ni des étiquettes directes de modèles de pointe.
Scores Codex et Gemma par catégorie
| Catégorie | Codex | Gemma |
|---|---|---|
| Métarecherche | 0,004 | 0,001 |
| Méta-épidémiologie (sens strict) | 0,000 | 0,000 |
| Méta-épidémiologie (sens large) | 0,000 | 0,000 |
| Bibliométrie | 0,000 | 0,000 |
| Études des sciences et des technologies | 0,000 | 0,000 |
| Communication savante | 0,000 | 0,000 |
| Science ouverte | 0,000 | 0,000 |
| Intégrité de la recherche | 0,000 | 0,000 |
| Charge utile insuffisante (le modèle a refusé de juger) | 0,000 | 0,000 |
Scores machine (provisoires)
Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.
Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.
score_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découleClassification
machine, non validéePrédiction automatique; un appel candidat d’une seule tête enseignante, pas un consensus.
Le détail, modèle par modèle et score par score, se trouve en fin de page sous « Comment cette classification a été obtenue ».