Technical Note: A novel quality assurance test to identify gantry angle inaccuracies in respiratory‐gated <scp>VMAT</scp> treatments
Notice bibliographique
Résumé
PURPOSE: During respiratory-gated volumetric-modulated arc therapy (VMAT), the radiation beam is turned off each time the target exits the gating window. At the same time, the gantry slows, stops, and rewinds before the beam is turned back on. A quality assurance (QA) test was developed to detect inaccuracies in the gantry angle position between beam-off and beam-on events during respiratory-gated VMAT. METHODS: 504 phantom mounted at isocenter. A homogeneous dose was delivered to the films through a 2 cm × 10 cm slit in the jaws using a respiratory-gated VMAT arc without the multileaf collimator. A periodic breathing cycle was used. Errors in gated delivery ranging from 0.5 to 5° were simulated by delivering nongated arcs with the same field size with over- and underlapping sections of 0.5-5°. The simulated errors were used to define QA levels to analyze the gated delivery. RESULTS: The QA test was capable of detecting errors as small as 0.5°. The test was delivered to three Varian TrueBeam™ linacs, and no gantry angle inaccuracies greater than or equal to 0.5° were detected on any of the films. CONCLUSIONS: A QA test capable of detecting gantry angle inaccuracies at beam-off and subsequent beam-on as small as 0.5° was developed and implemented for Varian TrueBeam™ linacs.
Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.
Comment cette classification a été obtenuedéplier
Prédiction machine sur la base complète
Imitation des enseignantsNi prévalence calibrée, ni vérité terrain. Validation humaine à venir. Le volet Gemma est une étiquette directe du modèle pour chaque travail de la base, lue sur la notice réduite au titre. Le volet Codex est un classifieur appris des 10 348 étiquettes directes de Codex et calibré sur les taux pondérés de l'échantillon; les champs sans appui suffisant ne portent aucun appel Codex. Le mode candidate est l'union des deux volets; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont pas des étiquettes humaines.
Scores du classifieur distillé par catégorie (deux têtes)
| Catégorie | Codex | Gemma |
|---|---|---|
| Métarecherche | 0,003 | 0,010 |
| Méta-épidémiologie (sens strict) | 0,001 | 0,000 |
| Méta-épidémiologie (sens large) | 0,000 | 0,000 |
| Bibliométrie | 0,001 | 0,000 |
| Études des sciences et des technologies | 0,000 | 0,000 |
| Communication savante | 0,001 | 0,001 |
| Science ouverte | 0,001 | 0,001 |
| Intégrité de la recherche | 0,001 | 0,000 |
| Charge utile insuffisante (le modèle a refusé de juger) | 0,002 | 0,001 |
Scores machine (provisoires)
Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.
Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.
score_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découleClassification
machine, non validéePrédiction automatique; un appel candidat d’une seule source (Gemma direct ou Codex distillé), pas un consensus.
Le détail, modèle par modèle et score par score, se trouve en fin de page sous « Comment cette classification a été obtenue ».