A 3D printed modular phantom for quality assurance of image‐guided small animal irradiators: Design, imaging experiments, and Monte Carlo simulations

2019· article· en· W2926818672 sur OpenAlexafffund
Dylan Y. Breitkreutz, Spencer Bialek, Anthony Kavanagh, Christopher Johnstone, Zach Rovner, Paul Tsouchlos, Thilakshan Kanesalingam, Magdalena Bazalova‐Carter

Notice bibliographique

RevueMedical Physics · 2019
Typearticle
Langueen
DomainePhysics and Astronomy
ThématiqueAdvanced Radiotherapy Techniques
Établissements canadiensUniversity of Victoria
Organismes subventionnairesNatural Sciences and Engineering Research Council of CanadaCanada Research ChairsCancer Research UK
Mots-clésImaging phantomImage qualityIsocenterImage resolutionMonte Carlo methodQuality assuranceComputer scienceNuclear medicineMaterials sciencePhysicsOpticsComputer visionMathematicsMedicine

Résumé

récupéré en direct d'OpenAlex

PURPOSE: The goal of this work was to develop and test a cylindrical tissue-equivalent quality assurance (QA) phantom for micro computed tomography (microCT) image-guided small animal irradiators that overcomes deficiencies of existing phantoms due to its mouse-like dimensions and composition. METHODS: The 8.6-cm-long and 2.4-cm-diameter phantom was three-dimensionally (3D) printed out of Somos NeXt plastic on a stereolithography (SLA) printer. The modular phantom consisted of four sections: (a) CT number evaluation section, (b) spatial resolution with slanted edge (for the assessment of longitudinal resolution) and targeting section, (c) spatial resolution with hole pattern (for the assessment of radial direction) section, and (d) uniformity and geometry section. A Python-based graphical user interface (GUI) was developed for automated analysis of microCT images and evaluated CT number consistency, longitudinal and radial modulation transfer function (MTF), image uniformity, noise, and geometric accuracy. The phantom was placed at the imaging isocenter and scanned with the small animal radiation research platform (SARRP) in the pancake geometry (long axis of the phantom perpendicular to the axis of rotation) with a variety of imaging protocols. Tube voltage was set to 60 and 70 kV, tube current was set to 0.5 and 1.2 mA, voxel size was set to 200 and 275 μm, imaging times of 1, 2, and 4 min were used, and frame rates of 6 and 12 frames per second (fps) were used. The phantom was also scanned in the standard (long axis of the phantom parallel to the axis of rotation) orientation. The quality of microCT images was analyzed and compared to recommendations presented in our previous work that was derived from a multi-institutional study. Additionally, a targeting accuracy test with a film placed in the phantom was performed. MicroCT imaging of the phantom was also simulated in a modified version of the EGSnrc/DOSXYZnrc code. Images of the resolution section with the hole pattern were acquired experimentally as well as simulated in both the pancake and the standard imaging geometries. The radial spatial resolution of the experimental and simulated images was evaluated and compared to experimental data. RESULTS: For the centered phantom images acquired in the pancake geometry, all imaging protocols passed the spatial resolution criterion in the radial direction (>1.5 lp/mm @ 0.2 MTF), the geometric accuracy criterion (<200 μm), and the noise criterion (<55 HU). Only the imaging protocol with 200-μm voxel size passed the criterion for spatial resolution in the longitudinal direction (>1.5 lp/mm @ 0.2 MTF). The 70-kV tube voltage dataset failed the bone CT number consistency test (<55 HU). Due to cupping artifacts, none of the imaging protocols passed the uniformity test of <55 HU. When the phantom was scanned in the standard imaging geometry, image uniformity and longitudinal MTF were satisfactory; however, the CT number consistency failed the recommended limit. A targeting accuracy of 282 and 251 μm along the x- and z-direction was observed. Monte Carlo simulations confirmed that the radial spatial resolution for images acquired in the pancake geometry was higher than the one acquired in the standard geometry. CONCLUSIONS: The new 3D-printed phantom presents a useful tool for microCT image analysis as it closely mimics a mouse. In order to image mouse-sized animals with acceptable image quality, the standard protocol with a 200-μm voxel size should be chosen and cupping artifacts need to be resolved.

Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.

Comment cette classification a été obtenuedéplier

Prédiction machine sur la base complète

Imitation des enseignants

Ni prévalence calibrée, ni vérité terrain. Validation humaine à venir. Le volet Gemma est une étiquette directe du modèle pour chaque travail de la base, lue sur la notice réduite au titre. Le volet Codex est un classifieur appris des 10 348 étiquettes directes de Codex et calibré sur les taux pondérés de l'échantillon; les champs sans appui suffisant ne portent aucun appel Codex. Le mode candidate est l'union des deux volets; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont pas des étiquettes humaines.

score de la tête « metaresearch » (Codex)0,001
score de la tête « metaresearch » (Gemma)0,001
Version: metacan-v3-hybrid-931329e0061cStatut de validation: machine_predicted_unvalidated
Catégories candidatesaucune
Catégories consensuellesaucune
DomaineSignal candidat: aucune · Signal consensuel: aucune
Devis d'étudeSignal candidat: Simulation ou modélisation · Signal consensuel: Simulation ou modélisation
GenreSignal candidat: Méthodes · Signal consensuel: Méthodes
Score de désaccord entre enseignants0,001
Score d'incertitude au seuil0,005

Scores du classifieur distillé par catégorie (deux têtes)

CatégorieCodexGemma
Métarecherche0,0010,001
Méta-épidémiologie (sens strict)0,0000,000
Méta-épidémiologie (sens large)0,0000,000
Bibliométrie0,0000,000
Études des sciences et des technologies0,0000,000
Communication savante0,0000,000
Science ouverte0,0010,000
Intégrité de la recherche0,0010,000
Charge utile insuffisante (le modèle a refusé de juger)0,0010,000

Scores machine (provisoires)

Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.

Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.

Tête enseignante Opus0,028
Tête enseignante GPT0,345
Écart entre enseignants0,316 · la distance entre les deux têtes enseignantes sur ce seul travail
Statut de validationscore_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découle

Classification

machine, non validée

Prédiction automatique; un appel candidat d’une seule source (Gemma direct ou Codex distillé), pas un consensus.

Les modèles n’ont appliqué aucune catégorie : rien dans la taxonomie ne correspondait à ce travail.
Devis d'étudeSimulation ou modélisation
Domainenon disponible
GenreMéthodes

Le détail, modèle par modèle et score par score, se trouve en fin de page sous « Comment cette classification a été obtenue ».

En bref

Citations9
Publié2019
Routes d'admission2
Résumé présentoui

Explorer davantage

Même revueMedical PhysicsMême sujetAdvanced Radiotherapy TechniquesTravaux en français237 207