Characterization of an x‐ray tube‐based ultrahigh dose‐rate system for in vitro irradiations
Notice bibliographique
Résumé
PURPOSE: To present an x-ray tube system capable of in vitro ultrahigh dose-rate (UHDR) irradiation of small < 0.3 mm samples and to characterize it by means of a plastic scintillation detector (PSD). METHODS AND MATERIALS: A conventional x-ray tube was modified for the delivery of short UHDR irradiations. A beam shutter system with a sample holder was designed and installed in a close proximity of an x-ray tube window to enable <1 s irradiations at UHDR. The dosimetry was performed with a small 0.5-mm long 0.5-mm in diameter PSD irradiated with 80, 100, and 120 kVp beams and beam currents of 1-37.5 mA. The PSD signal was recorded at frame rates of 20 and 50 fps for shutter exposure between 100 and 1125 ms. Irradiation reproducibility was studied with the PSD. The x-ray tube irradiation setup was modeled with Monte Carlo (MC) and dose on a surface of a phantom was also measured with films. The effect of dose delivery uncertainty to 300-μm spheroids due to positioning and spheroid size was evaluated. RESULTS: MC simulations showed good agreement with PSD measurements acquired at both frame rates of 20 and 50 fps in terms of beam temporal profile. PSD-measured dose exhibited excellent linearity as a function of instantaneous dose rate from 3.1 to 118.0 Gy/s as well as shutter exposure time from 100 and 1125 ms for all investigated beam energies. PSD absorbed dose for the 80, 100, and 120 kVp beams agreed with MC simulations to within 5%. The total delivered doses ranged from 0.4 Gy for a 1-mA, 80 kVp beam, and 100 ms shutter exposure to 166.9 Gy for a 37.5-mA, 80 kVp beam, and a 1125 ms exposure. PSD irradiation reproducibility was < 0.5%. Simulated and measured dose fall off agreed and it was steep along the axis of the shutter slit (1%/0.1 mm) and with depth (2%/0.1 mm at 1-mm depth). Spheroid positioning uncertainty of 300 μm resulted in dose difference of < 3% for x and y shifts but up to 7% uncertainty for a z-shift parallel to the beam axis. A 16% difference in spheroid size resulted in <5% dose difference in spheroid absorbed dose. CONCLUSIONS: We have presented a cost-effective x-ray tube-based system with a beam shutter designed for in vitro UHDR delivery and reaching dose rates of up to 118.0 Gy/s. The described shutter system can be easily implemented at other institutions, which might enable new researchers to investigate the radiobiology of UHDR irradiations in vitro.
Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.
Comment cette classification a été obtenuedéplier
Prédiction distillée sur la base complète
Imitation des enseignantsNi prévalence calibrée, ni vérité terrain. Validation humaine à venir. Apprise à partir de 10 348 étiquettes directes de Codex et de 10 348 étiquettes directes de Gemma. Le mode candidate est l'union des têtes enseignantes seuillées; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont ni des étiquettes humaines ni des étiquettes directes de modèles de pointe.
Scores Codex et Gemma par catégorie
| Catégorie | Codex | Gemma |
|---|---|---|
| Métarecherche | 0,000 | 0,000 |
| Méta-épidémiologie (sens strict) | 0,000 | 0,000 |
| Méta-épidémiologie (sens large) | 0,000 | 0,000 |
| Bibliométrie | 0,000 | 0,000 |
| Études des sciences et des technologies | 0,000 | 0,000 |
| Communication savante | 0,000 | 0,000 |
| Science ouverte | 0,000 | 0,000 |
| Intégrité de la recherche | 0,000 | 0,000 |
| Charge utile insuffisante (le modèle a refusé de juger) | 0,000 | 0,000 |
Scores machine (provisoires)
Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.
Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.
score_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découleClassification
machine, non validéePrédiction automatique; un appel candidat d’une seule tête enseignante, pas un consensus.
Le détail, modèle par modèle et score par score, se trouve en fin de page sous « Comment cette classification a été obtenue ».