Population‐level anemia prevalence rates may be rendered inaccurate when hemoglobin is measured in pooled capillary blood or with the <scp>HemoCue</scp>® 301 device
Notice bibliographique
Résumé
Anemia, defined as a low hemoglobin concentration, impacts ~one third of women of reproductive age (WRA) and children globally, with the highest rates in low-resource settings.1 Venous blood measured with hematology autoanalyzers are currently recommended by the World Health Organization for hemoglobin measurement; however, this is not always feasible. Portable hemoglobinometers, such as the HemoCue® (Angelholm, Sweden), are often used in large field studies or anemia surveys due to their ease of use and their ability to assess hemoglobin in a single-drop of capillary blood.1 However, there is growing concern that single-drop capillary blood introduces too much variability (random error) to hemoglobin estimates, which leads to inaccurate estimation of population-level anemia prevalence.2, 3 Pooled capillary blood offers a possible alternate blood source to single-drop capillary blood, but there is limited and mixed evidence on whether pooled capillary specimens have the same inherent issues as single-drop capillary blood4; thus, more research in this area is urgently required. The recent HEmoglobin MEeasurement (HEME) multicountry study5 evaluated repeated hemoglobin measurements using a variety of analytical methods and blood sources. Using data from our Cambodian HEME study site, we had the unique opportunity to evaluate the use of pooled capillary blood for hemoglobin measurement, and offer novel commentary for its use as compared with gold standard methods. In brief, pooled capillary blood was collected using contact-activated lancets (BD Medical, USA). After wiping away the first drop of blood, a 1 mL EDTA vacutainer (Greiner Bio-One, Austria) was placed at the base of the puncture site to collect the pooled sample (~250 μL blood; ~8–15 drops) within 2 min. Venous blood was collected concurrently in a 2 mL EDTA vacutainer (BD Medical, USA). Hemoglobin concentrations were measured with three HemoCue® models (Hb 201+, 301, and 801) using pooled capillary blood and venous blood; venous blood was further evaluated in a hematology autoanalyzer (Sysmex XN-1000, Sysmex Corp., Japan). All HemoCue® devices performed well by manufacturer standards and reference materials (HemoTrol® QC solution) tested within acceptable ranges. We calculated the mean difference (95% CI) in hemoglobin measurements across methods. Repeated measures from a total of 36 participants (18 WRA and 18 children aged 12–59 months) were included in this assessment, and compared with a repeated measures ANOVA (with Bonferroni correction for any multiple comparisons); see Table 1. Across WRA and children (data pooled), mean (95% CI) hemoglobin concentrations using pooled capillary blood measured in the HemoCue® models (201+, 301, and 801) were 3.5 (1.5, 5.5) g/L, 9.2 (7.4, 11) g/L, and 5.1 (2.6, 7.5) g/L higher, respectively, as compared with the gold standard method (venous blood via the autoanalyzer) (Figure 1). Separate data for WRA and children, and the corresponding anemia prevalence rates for each group are also presented in Table 1. This overestimation of hemoglobin concentrations driven by the use of pooled capillary blood is further evidenced in Figure 2; across WRA and children (data pooled), mean (95% CI) hemoglobin concentrations using pooled capillary blood measured in the HemoCue® models (201+, 301, and 801) were 4.4 (1.6, 7.1) g/L, 1.7 (−0.8, 4.2) g/L, and 2.8 (−0.05, 5.7) g/L higher, respectively, as compared with venous blood from the same individuals repeated in the same HemoCue® models. Ultimately, the use of pooled capillary blood resulted in a systematic overestimation of hemoglobin concentrations and concurrent underestimation of the prevalence of population-level anemia; the most predominant underestimation of anemia found via use of the HemoCue® 301 device. These findings suggest that population-level anemia prevalence rates may be rendered inaccurate when hemoglobin is measured in pooled capillary blood or with the HemoCue® 301 device. Our findings are aligned with much of the previous literature, which reports that differences in analytical methods can substantially influence hemoglobin results.6, 7 This may be driven by many factors, such as operator-experience, humidity, and the time between collection and measurement8-10; these factors typically vary across protocols and settings. Of note, blood collection and quantification techniques in the current study were performed as per a detailed protocol that was developed by anemia experts of the larger HEME study,5 which aimed to minimize potential biases (e.g., using only one comprehensively-trained operator, detailed specimen handling, time between collection, and analysis <1 min). However, this study reinforces caution against the comparison of hemoglobin estimates or population-level anemia prevalence rates across surveys or timepoints when different analytical approaches have been used. Furthermore, a crucial gap in current literature is that most hemoglobin method-comparison studies cannot tease out whether differences across methods are due to the blood source or the analytical device; we address this gap in our investigation. We found that hemoglobin concentrations measured with pooled capillary blood were systematically higher than when measured with venous blood via the autoanalyzer or with the same HemoCue® devices (suggesting that differences were due to the blood source). However, we note some uncertainty as to whether this was a result of biological differences in the blood source or simply due to the higher variability in capillary blood as compared with venous blood (as evidenced by other studies2, 3, 11). Regardless, we caution against use of pooled capillary blood specimens due to these apparent inaccuracies. Finally, while overestimation was observed across all three HemoCue® devices, the greatest inaccuracies in estimation of anemia prevalence was observed in the 301 device. Others have reported similar issues of overestimating hemoglobin concentration with the HemoCue® 301 device,10 which is particularly problematic as it is the most commonly used model for use in National Demographic and Health Surveys. If 301 devices are consistently and systematically overestimating hemoglobin concentrations, this would result in a global underestimation of anemia prevalence. It has recently been suggested that this type of systematic bias could potentially be adjusted for with the validation of each HemoCue® device before the start of a survey2; however, there is no global consensus on this approach and whether it is appropriate for all settings and populations. Further, this approach poses challenges if the directionality of the observed systematic bias differs across the range of low and high hemoglobin concentrations, as we have reported in a previous study.12 Overall, these finding are timely and of critical relevance considering the recent release of the 2024 World Health Organization guidelines on hemoglobin measurement and the revised thresholds for anemia diagnosis (https://www.who.int/publications/i/item/9789240088542). Countries on the verge of implementing DHS surveys or large anemia studies may be conflicted on what blood source and analytical device to use, given a lack of evidence or guidance regarding the use of pooled capillary blood specimens. Further research is needed to determine if procedures for the collection of pooled capillary blood could be further optimized to reduce variability and overestimation. At this time, we echo sentiments to follow the gold standard methods, which is the collection of venous blood for hemoglobin measurement with use of an automated hematology analyzer. CDK and HK conceptualized the investigation; KMC and BAW assisted with field implementation, data and blood specimen collection, and formal analysis; AC assisted with field implementation and blood specimen collection; HK supervised research staff in the field; KMC, BAW, and CDK contributed to writing the original draft; HK and AC contributed to review and editing. The authors have accepted responsibility for the entire content of this manuscript and approved its submission. We thank Ngik Rem for his assistance with training and study implementation in Cambodia. The study was funded by the U.S. Agency for International Development (USAID) under the terms of contract 7200AA18C00070 awarded to JSI Research & Training Institute. The contents are the responsibility of the authors, and do not necessarily reflect the views of USAID or the US Government. CDK is supported by a Michael Smith Foundation for Health Research Scholar Award and holds a Canada Research Chair in Micronutrients and Human Health. BAW was supported by a Michael Smith Foundation for Health Research (#180216). The authors declare no conflict of interest. Data is available upon reasonable request to the principal investigator (CD Karakochuk).
Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.
Comment cette classification a été obtenuedéplier
Prédiction machine sur la base complète
Imitation des enseignantsNi prévalence calibrée, ni vérité terrain. Validation humaine à venir. Le volet Gemma est une étiquette directe du modèle pour chaque travail de la base, lue sur la notice réduite au titre. Le volet Codex est un classifieur appris des 10 348 étiquettes directes de Codex et calibré sur les taux pondérés de l'échantillon; les champs sans appui suffisant ne portent aucun appel Codex. Le mode candidate est l'union des deux volets; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont pas des étiquettes humaines.
Scores du classifieur distillé par catégorie (deux têtes)
| Catégorie | Codex | Gemma |
|---|---|---|
| Métarecherche | 0,002 | 0,014 |
| Méta-épidémiologie (sens strict) | 0,000 | 0,000 |
| Méta-épidémiologie (sens large) | 0,000 | 0,000 |
| Bibliométrie | 0,001 | 0,001 |
| Études des sciences et des technologies | 0,001 | 0,001 |
| Communication savante | 0,001 | 0,001 |
| Science ouverte | 0,001 | 0,000 |
| Intégrité de la recherche | 0,009 | 0,006 |
| Charge utile insuffisante (le modèle a refusé de juger) | 0,011 | 0,008 |
Scores machine (provisoires)
Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.
Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.
score_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découleClassification
machine, non validéePrédiction automatique; un appel candidat d’une seule source (Gemma direct ou Codex distillé), pas un consensus.
Le détail, modèle par modèle et score par score, se trouve en fin de page sous « Comment cette classification a été obtenue ».