SU‐F‐18C‐10: Clinical Implementation of Radiochromic Film Based CTDI Measurements

2014· article· en· W1970643799 on OpenAlexaffabout
C. Quintero, Nada Tomic, F DeBlois, Jan Seuntjens, Slobodan Dević

Bibliographic record

VenueMedical Physics · 2014
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsMcGill UniversityMontreal General HospitalJewish General Hospital
Fundersnot available
KeywordsKermaScannerMaterials scienceDosimetryImaging phantomNuclear medicineReproducibilityCalibrationIonization chamberQuality assuranceImage qualityOpticsPhysicsMedicineComputer scienceMathematics

Abstract

fetched live from OpenAlex

Purpose: To evaluate accuracy and reproducibility of a radiochromic film‐based protocol to measure computer tomography dose index (CTDI) as a part of annual QA on CT scanners and kV CBCT systems attached to linear accelerators. Methods: Energy dependence of Gafchromic XR‐QA2(R) film model was tested over imaging beam qualities (50 – 140 kVp). Film pieces were irradiated in air to known values of air kerma in air (up to 10 cGy). Change in film reflectance was determined with an in‐house written code using images produced by a flatbed document scanner. Calibration curves for each beam quality were created, and film responses for same air‐kerma values were compared.Sets of film strips were placed into holes of a CTDI phantom and irradiated for several clinical scanning protocols on CT‐simulators and CBCT systems. Film reflectance change was converted into dose to water and used to calculate CTDIvol values. Measured CTDIvol values were compared to tabulated CTDIvol values. Results: Average variations of ±5.2% in the mean film reflectance change were observed in the energy range of 80 to 140 keV, and 11.1% between 50 and 140 keV. The averaged measured CTDI values presented a mean variation for the same machine and protocol of 2.6%. However, measured CTDI values were in average 10% lower than tabulated CTDI values for CT‐simulators, and 44% higher for CBCT systems. Conclusion: We found that in relatively broad range of beam qualities used in diagnostic radiology variation of film response is within ±5% resulting in ±15% systematic error in dose estimates if a single calibration curve is used. Relatively large discrepancy between measured and tabulated CTDI values for different protocols and imaging systems used within radiotherapy department strongly support the trend towards replacing CTDI value with equilibrium dose measurement in the center of cylindrical phantom as suggested by TG‐111. This work was supported by the Natural Sciences and Engineering Research Council of Canada, Contract No. 386009; Partial support by CREATE, Medical Physics Research Training Network grant of the Natural Science and Engineering Research Council, Contract No. 432290.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.003
metaresearch head score (Gemma)0.007
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.003
Threshold uncertainty score0.017

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0030.007
Meta-epidemiology (narrow)0.0010.001
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0010.001
Science and technology studies0.0000.001
Scholarly communication0.0010.000
Open science0.0010.000
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0020.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.036
GPT teacher head0.384
Teacher spread0.348 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2014
Admission routes2
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced Radiotherapy Techniques→French-language works237,207→