SU-FF-T-209: Electron Arc Therapy with An Elekta SL-25 10 MeV Beam Using a Dedicated Short Applicator

2006· article· en· W1980068302 on OpenAlexaff
C Duchesne, M. Mondat, W Wierzbicki

Bibliographic record

VenueMedical Physics · 2006
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsHôpital Maisonneuve-Rosemont
Fundersnot available
KeywordsImaging phantomIsocenterBeam (structure)Monitor unitPhysicsLinear particle acceleratorRADIUSOpticsCathode rayArc (geometry)Nuclear medicineMathematicsElectronGeometryNuclear physicsComputer science

Abstract

fetched live from OpenAlex

Purpose: To implement the electron arc therapy with an Elekta SL-25 10 Mev beam using a dedicated short applicator. Method and Materials:. All measurements were done on an Elekta SL-25 with a 10 MeV electron beam. Radial PDD were characterized for 4 different cylindrical acrylic phantoms using TLDs and Kodak X-Omat-V radiographic films The phantom radius (di) ranged between 5.1 and 15.2 cm, the total arc angles (α) varied between 60° and 160° and the number of monitor units (MU) per degree between 0.5 and 7. The field with at isocenter (w) was varied between 3.7 and 9.8 cm. Based on this data bank, an analytical model was developed for monitor unit (MU) calculation. This model estimates arc beam output at the depth of maximum dose (dmax) as a function of di and α for a given w. A verification of the model precision was done with a new cylindrical phantom. Results: Curve fitting of the complete set of beam output data with w fixed at 7.3 cm was done with an asymptotic relationship between the dose rate at dmax and the inverse square of di. The dependence of the beam output on α was introduced by assuming an explicit function of α for each parameter of the model. Results show that the calculated beam output data is a good approximation for all measured data for all phantoms and arc angles: 89% of the calculated values are within ± 3% of the measured ones and all points are within a ± 5% error range. Conclusion: From our results, clinical implementation of the electron arc therapy technique is possible and should be facilitated by the use of our predictive model in the treatment planning process.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.003
Threshold uncertainty score0.010

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0010.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0030.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.010
GPT teacher head0.286
Teacher spread0.276 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2006
Admission routes1
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced Radiotherapy Techniques→French-language works237,207→