MétaCan
Menu
Back to cohort
Record W1998455099 · doi:10.1118/1.3539575

The effects of compensator and imaging geometry on the distribution of x‐ray scatter in CBCT

2011· article· en· W1998455099 on OpenAlexaff
Gregory Bootsma, Frank Verhaegen, David A. Jaffray

Bibliographic record

VenueMedical Physics · 2011
Typearticle
Languageen
FieldMedicine
TopicDigital Radiography and Breast Imaging
Canadian institutionsMcGill UniversityPrincess Margaret Cancer CentreUniversity of TorontoOntario Institute for Cancer Research
FundersElekta
KeywordsCone beam computed tomographyMonte Carlo methodDetectorOpticsFlat panel detectorPhysicsProjection (relational algebra)Field of viewScatteringMedical imagingComputed tomographyMathematicsComputer scienceArtificial intelligenceMedicineAlgorithmStatistics

Abstract

fetched live from OpenAlex

PURPOSE: X-ray scatter contributes significantly to image degradation in cone-beam CT (CBCT) reconstructed images in the form of CT number inaccuracy image artifacts and loss of contrast. The need for an understanding of the relationship between the scatter distribution and common imaging parameters (cone angle, air gap, filtration, object size) is an essential element to developing methods for efficiently correcting for the effects of scatter in CBCT. The first objective of this study is to validate the scatter distributions calculated using a CBCT Monte Carlo (MC) model against measured scatter estimates. The second objective is to use the CBCT MC model to investigate the effects of common imaging parameters and bowtie compensators on the resulting scatter distribution. METHODS: This investigation employs the use of a CBCT MC model, developed using the EGSnrc code, to simulate the primary and scatter fluence arriving at the detector. The simulation is validated against projection images, scatter-to-open field ratio (SOR), and scatter-to-primary ratio (SPR) measurements taken using a bench-top CBCT system. The CBCT MC model is used to simulate the scatter distribution arriving at the detector for different cone angles {1.4 degrees, 2.8 degrees, 5.7 degrees, and 11.3 degrees}, source-to-axis distances (SADs) {50, 75 and 100 cm}, and axis-to-detector distances (ADDs) {9, 18, 30, 44, 56 cm} for both a 16.4 and 30.6 cm diameter water cylinder. The effects of different bowtie filters are also simulated using the CBCT MC model for the aforementioned cylinder sizes. RESULTS: Profiles of the simulated and measured projection images agree within 6%. The measurements of the SOR and SPR, taken using beam stop techniques, show good agreement with the simulated results. Limitations of current beam stop techniques in estimating scatter profiles of objects with varying thickness are also demonstrated. A functional relationship between scatter (SOR, SPR) and air gap, cone angle is reported. The bowtie filter was found to have the beneficial effect of decreasing the magnitude of scatter in the projection images (SPR decreased by 56%) as well as altering the spatial distribution of the scatter. CONCLUSIONS: The CBCT MC model accurately simulates scatter and primary fluences in the CBCT imaging components and geometry. The CBCT MC model provides a useful tool in investigating the effects of varying CBCT imaging parameters on the scatter distribution. Increasing the air gap, decreasing the cone angle, and the use of bowtie filtration were all found to be effective ways to minimize scatter in CBCT. The bowtie filter was particularly effective in both minimizing the magnitude and modifying the spatial distribution of the scattered photons. This observation directs further research in optimizing bowtie filter design in an effort to minimize scatter induced artifacts.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: Observational
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.475
Threshold uncertainty score0.173

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.008
GPT teacher head0.227
Teacher spread0.218 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations58
Published2011
Admission routes1
Has abstractyes

Explore more

Same venueMedical PhysicsSame topicDigital Radiography and Breast ImagingFrench-language works237,207