Development and Application of a Needle Trap Device for Time-Weighted Average Diffusive Sampling

2008· article· en· W2025557888 on OpenAlexafffund
Ying Gong, In‐Yong Eom, Dawei Lou, D. Hein, Janusz Pawliszyn

Bibliographic record

VenueAnalytical Chemistry · 2008
Typearticle
Languageen
FieldChemistry
TopicAnalytical chemistry methods development
Canadian institutionsUniversity of Waterloo
FundersNatural Sciences and Engineering Research Council of CanadaU.S. Environmental Protection Agency
KeywordsChemistrySorbentBTEXEthylbenzeneChromatographyThermal desorptionAdsorptionAnalytical Chemistry (journal)Sampling (signal processing)DesorptionDiffusionXyleneSample preparationTolueneOrganic chemistryThermodynamics

Abstract

fetched live from OpenAlex

A simple, cost-effective analysis combining solventless extraction, thermal desorption, and determination of volatile organic compounds (VOCs) was developed and validated. A needle trap device (NTD) packed with the sorbent Carboxen1000 was used as a time-weighted average (TWA) diffusive sampler to collect target compounds by molecular diffusion and adsorption to the packed sorbent. This process can be described with derivations of Fick's first law of diffusion, which expresses the relation between the TWA concentrations to which the passive sampler is exposed and the mass of analytes adsorbed to the packed sorbent in the sampler. The effects of experimental factors such as temperature, pressure, humidity, and face velocity were taken into account in applying diffusive sampling under nonideal conditions. This study demonstrates that NTD is effective for air analysis of benzene, toluene, ethylbenzene, and o-xylene (BTEX), due to the good adsorption/desorption quality of Carboxen 1000 and to the special geometric shape of the needle with a small cross section avoiding the need for calibration. Storage tests showed good storage stability for BTEX. Verification of the theoretical model showed good agreement between theoretical and experimental sampling rates. Method validation done against NIOSH method 1501, SPME, and NTD active sampling revealed good agreement between those methods. Automated NTD sample introduction to a gas chromatograph facilitates the use of this technology for industrial hygiene applications.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesMeta-epidemiology (narrow)
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.026
Threshold uncertainty score1.000

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0010.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.035
GPT teacher head0.294
Teacher spread0.258 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations73
Published2008
Admission routes2
Has abstractyes

Explore more

Same venueAnalytical ChemistrySame topicAnalytical chemistry methods developmentFrench-language works237,207