Sci-Fri AM(2): Brachy-02: Realizing Absorbed Dose Directly for HDR Ir-192 Brachytherapy: Water Calorimetry and comparison to Ion Chamber, Gafchromic Film, and TG-43

2009· article· en· W2030110143 on OpenAlexaff
Arman Sarfehnia, Jan Seuntjens

Bibliographic record

VenueMedical Physics · 2009
Typearticle
Languageen
FieldMedicine
TopicRadiation Therapy and Dosimetry
Canadian institutionsMcGill University
Fundersnot available
KeywordsDosimetryIonization chamberCalorimetryMaterials scienceAbsorbed doseBrachytherapyDosimeterNuclear medicineRadiochemistryAnalytical Chemistry (journal)IonizationPhysicsChemistryIonMedicineRadiation therapyRadiology

Abstract

fetched live from OpenAlex

A new standard based on direct measurement of absolute absorbed dose to water in HDR brachytherapy using water calorimetry is discussed. The calorimetry results are compared with the results from the in-water measurement of absorbed dose using ionization chamber and EBT Gafchromic film reference dosimetry, as well as TG-43. The water calorimetry measurements were performed using a spring-loaded catheter holder system that mounts onto a parallel-plate calorimetric vessel. The COMSOL MULTIPHYSICS™ software was used to perform the numerical heat-transport calculations. The calorimetric measurements were performed with three different sources with nominal ranging from 31600–36940 U (i.e. 7.6–9.2 Ci activity) over a one-year period. A source-to-detector separation ranging between 51–67 mm, and an irradiation time ranging between 200–300 s were used. Based on water calorimetry, an average dose rate to water of 361±11 μGy/(h.U) at mm was measured. Chamber reference dosimetry resulted in a dose rate measurement of 366±9 μGy/(h.U), while EBT Gafchromic film reference dosimetry resulted in 368±11 μGy/(h.U). From TG-43 protocol, we obtain a value of 364±15 μGy/(h.U), a 0.8% difference from water calorimetry. All Results agree well to within uncertainty. Our work paves the way to successful primary absorbed dose determination for using calorimetric techniques.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.000
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.006
Threshold uncertainty score0.019

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0010.001
Science and technology studies0.0000.001
Scholarly communication0.0010.000
Open science0.0000.001
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0060.002

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.021
GPT teacher head0.315
Teacher spread0.294 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2009
Admission routes1
Has abstractyes

Explore more