MétaCan
Menu
Back to cohort
Record W2039769504 · doi:10.1120/jacmp.v3i4.2553

Comparison of monitor unit calculations performed with a 3D computerized planning system and independent “hand” calculations: Results of three years clinical experience

2002· article· en· W2039769504 on OpenAlexaff
Jackson Chan, David Russell, Victor Peters, Thomas J. Farrell

Bibliographic record

VenueJournal of Applied Clinical Medical Physics · 2002
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsHamilton Regional Laboratory Medicine Program
Fundersnot available
KeywordsMonitor unitRadiation treatment planningUnit (ring theory)Computer scienceMathematicsNuclear medicineMedical physicsMedicineSurgery

Abstract

fetched live from OpenAlex

A comparison of the monitor unit calculations of a commercial 3D computerized treatment planning system (TPS) with "hand" calculations from lookup tables was made for a large number of clinical cases (greater than 13 500 treatment fields). Differences were analyzed by treatment site for prostate, rectum, cranium, and breast. The 3D TPS monitor unit calculation was systematically higher than the "hand" calculation by an amount that depended on the complexity of the treatment geometry. For simple geometries the mean difference was 1% and was as high as 3% for more complicated geometries. The higher value was attributed to an accumulation of differences introduced by multiple factors in the monitor unit calculation. Careful attention to factors such as patient contour could reduce the mean difference. "Hand" calculations were shown to be an accurate and useful tool for verification of TPS monitor unit calculations.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: Observational
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.414
Threshold uncertainty score0.550

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0010.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0000.000
Science and technology studies0.0000.001
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.001
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.087
GPT teacher head0.414
Teacher spread0.326 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations17
Published2002
Admission routes1
Has abstractyes

Explore more

Same venueJournal of Applied Clinical Medical PhysicsSame topicAdvanced Radiotherapy TechniquesFrench-language works237,207