MétaCan
Menu
Back to cohort
Record W2067297320 · doi:10.1120/jacmp.v6i2.2083

Validation of physics improvements for IMRT with a commercial treatment‐planning system

2005· article· en· W2067297320 on OpenAlex

Why this work is in the frame

A frame that forgets how it found something cannot be audited. These are the routes that admitted this work.

affAt least one author lists a Canadian institution in the pinned OpenAlex snapshot.

Bibliographic record

VenueJournal of Applied Clinical Medical Physics · 2005
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsSaskatoon Medical ImagingSaskatchewan Cancer Agency
Fundersnot available
KeywordsPinnacleMultileaf collimatorDosimeterRadiation treatment planningIonization chamberCollimatorSoftwareSensitivity (control systems)OpticsNuclear medicineMaterials sciencePhysicsMedical physicsBeam (structure)Computer scienceRadiation therapyLinear particle acceleratorRadiationEngineeringMedicineElectronic engineeringSurgery

Abstract

fetched live from OpenAlex

A new Pinnacle 3D treatment-planning system software release has recently become available (v7.4, Philips Radiation Oncology Systems, Milpitas, CA), which supports modeling of rounded multileaf collimator (MLC) leaf ends; it also includes a number of other software enhancements intended to improve the overall dose calculation accuracy. In this report, we provide a general discussion of the dose calculation algorithm and new beam-modeling parameters. The accuracy of a diode dosimeter was established for measurement of MLC-shaped beam profiles required by the new software version by comparison with film and ion chamber measurements in various regions of the field. The results suggest that a suitable diode or other small volume dosimeter with appropriate energy sensitivity should be used to obtain profiles for commissioning the planning system. Film should be used with caution, especially for larger field profile measurements. The dose calculation algorithm and modeling parameters chosen were validated through various test field measurements including a bar pattern, a strip pattern, and a clinical head and neck IMRT field. For the bar and strip patterns, the agreement between Pinnacle calculations and diode measurements was generally very good. These tests were helpful in establishing the new model parameter values, especially tongue-and-groove width, additional interleaf leakage, rounded leaf tip radius, and MLC transmission. For the clinical head and neck field, the comparison between Pinnacle and film measurements showed regions of approximately 2 cGy under- or overdose. However, the Pinnacle calculations agreed with diode measurements at all points to within 1 cGy or 1% of the maximum dose for the field (67 cGy). The greatest discrepancy between film and diode measurements for the clinical field (maximum of 2.8%) occurred in low-dose regions in the central part of the field. The disagreement may be due to the overresponse of film to scattered radiation in the low-dose regions, which have significant shielding by the MLCs.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Other design · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: none
Teacher disagreement score0.958
Threshold uncertainty score0.568

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.036
GPT teacher head0.383
Teacher spread0.347 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it