MétaCan
Menu
Back to cohort
Record W2091693485 · doi:10.1118/1.3692177

Biological consequences of MLC calibration errors in IMRT delivery and QA

2012· article· en· W2091693485 on OpenAlexaff
Vitali Moiseenko, Vincent Lapointe, Kerry James, Lingshu Yin, Mitchell Liu, Todd Pawlicki

Bibliographic record

VenueMedical Physics · 2012
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsBC Cancer Agency
Fundersnot available
KeywordsNuclear medicineMultileaf collimatorMedicineIonization chamberDosimetryImaging phantomHead and neckRadiation treatment planningRadiation therapyCalibrationProstateRadiologyMathematicsSurgeryPhysics

Abstract

fetched live from OpenAlex

PURPOSE: The purpose of this work is threefold: (1) to explore biological consequences of the multileaf collimator (MLC) calibration errors in intensity modulated radiotherapy (IMRT) of prostate and head and neck cancers, (2) to determine levels of planning target volume (PTV) and normal tissue under- or overdose flagged with clinically used QA action limits, and (3) to provide biologically based input for MLC QA and IMRT QA action limits. METHODS: Ten consecutive prostate IMRT cases and ten consecutive head and neck IMRT cases were used. Systematic MLC offsets (i.e., calibration error) were introduced for each control point of the plan separately for X1 and X2 leaf banks. Offsets were from - 2 to 2 mm with a 0.5 mm increment. The modified files were imported into the planning system for forward dose recalculation. The original plan served as the reference. The generalized equivalent uniform dose (gEUD) was used as the biological index for the targets, rectum, parotid glands, brainstem, and spinal cord. Each plan was recalculated on a CT scan of a 27 cm diameter cylindrical phantom with a contoured 0.6 cc ion chamber. Dose to ion chamber and 3D gamma analysis were compared to the reference plan. QA pass criteria: (1) at least 95% of voxels with a dose cutoff of 50% of maximum dose have to pass at 3 mm/3% and (2) dose to chamber within 2% of the reference dose. RESULTS: For prostate cases, differences in PTV and rectum gEUD greater than 2% were identified. However, a larger proportion of plans leading to greater than 2% difference in prostate PTV gEUD passed the ion chamber QA but not 3D gamma QA. A similar trend was found for the rectum gEUD. For head and neck IMRT, the QA pass criteria flagged plans leading to greater than 4% differences in PTV gEUD and greater than 5% differences in the maximum dose to brainstem. If pass criteria were relaxed to 90% for gamma and 3% for ion chamber QA, plans leading to a 5% difference in PTV gEUD and a 5%-8% difference in brainstem maximum dose would likely pass IMRT QA. A larger proportion of head and neck plans with greater than 2% PTV gEUD difference passed 3D gamma QA compared to ion chamber QA. CONCLUSIONS: For low modulation plans, there is a better chance to catch MLC calibration errors with 3D gamma QA rather than ion chamber QA. Conversely, for high modulation plans, there is a better chance to catch MLC calibration errors with ion chamber QA rather than with 3D gamma QA. Ion chamber and 3D gamma analysis IMRT QA can detect greater than 2% change in gEUD for PTVs and critical structures for low modulation treatment plans. For high modulation treatment plans, ion chamber and 3D gamma analysis can detect greater than 2% change in gEUD for PTVs and a 5% change in critical structure gEUD since either QA methods passes the QA criteria. For gEUD changes less than those listed above, either QA method has the same proportion of passing rate.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.804
Threshold uncertainty score0.232

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.022
GPT teacher head0.295
Teacher spread0.273 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations20
Published2012
Admission routes1
Has abstractyes

Explore more

Same venueMedical PhysicsSame topicAdvanced Radiotherapy TechniquesFrench-language works237,207