MétaCan
Menu
Back to cohort
Record W2095573140 · doi:10.1117/12.795542

Development of ultrafast laser-based x-ray in-vivo phase-contrast micro-CT beamline for biomedical applications at Advanced Laser Light Source (ALLS)

2008· article· en· W2095573140 on OpenAlexaff
Russell Kincaid, Andrzej Król, S. Fourmaux, Jean-Claude Kieffer, Cristina Şerbănescu, Marina Servol, Levon Vogelsang, S. Wilkins, A. W. Stevenson, Yakov I. Nesterets, Edward D. Lipson, Hongwei Ye, A. Pogany

Bibliographic record

VenueProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2008
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced X-ray Imaging Techniques
Canadian institutionsUniversité du Québec
FundersNational Institute of Biomedical Imaging and BioengineeringSyracuse University
KeywordsOpticsLaserBeamlinePhase-contrast imagingLaser ablationPhysicsUltrashort pulseMaterials sciencePhase (matter)Image resolutionBeam (structure)Phase contrast microscopy

Abstract

fetched live from OpenAlex

We are developing and exploring the imaging performance of, an in vivo, in-line holography, x-ray phase-contrast, micro-CT system with an ultrafast laser-based x-ray (ULX) source. By testing and refining our system, and by performing computer simulations, we plan to improve system performance in terms of contrast resolution and multi-energy imaging to a level beyond what can be obtained using a conventional microfocal x-ray tube. Initial CT projection sets at single energy (Mo K(alpha) and K(beta) lines) were acquired in the Fresnel regime and reconstructed for phantoms and a euthanized mouse. We also performed computer simulations of phase-contrast micro-CT scans for low-contrast, soft-tissue, tumor imaging. We determined that, in order to perform a phase-contrast, complete micro-CT scan using ULX, the following conditions must be met: (i) the x-ray source needs to be stable during the scan; (ii) the laser focal spot size needs to be less than 10 mum for source-to-object distance greater than 30 cm; (iii) the laser light intensity on the target needs to be in the range of 5 x 10(17) to 5 x 10(19) W/cm(2); (iv) the ablation protection system needs to allow uninterrupted scans; (v) the laser light focusing on the target needs to remain accurate during the entire scan; (vi) a fresh surface of the target must be exposed to consecutive laser shots during the entire scan; (vii) the effective detector element size must be less than 12 mum. Based on the results obtained in this research project, we anticipate that the new 10 Hz, 200 TW laser with 50 W average power that is being commissioned at ALLS will allow us practical implementation of in vivo x-ray phase-contrast micro-CT.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesMeta-epidemiology (narrow)
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.467
Threshold uncertainty score1.000

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0010.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.009
GPT teacher head0.244
Teacher spread0.236 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations5
Published2008
Admission routes1
Has abstractyes

Explore more

Same venueProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIESame topicAdvanced X-ray Imaging TechniquesFrench-language works237,207