Record W2098216708 · doi:10.3109/02713683.2012.757324

Optimization of a Fluorescence-based Lysozyme Activity Assay for Contact Lens Studies

2013· article· en· W2098216708 on OpenAlexafffund
Alan Ng, Miriam Heynen, Doerte Luensmann, Lakshman N. Subbaraman, Lyndon Jones

Bibliographic record

VenueCurrent Eye Research · 2013
Typearticle
Languageen
FieldMedicine
TopicOcular Surface and Contact Lens
Canadian institutionsUniversity of Waterloo
FundersJohnson and Johnson Vision CareNatural Sciences and Engineering Research Council of CanadaAbbott Medical OpticsCanadian Optometric Education Trust FundCooperVision
KeywordsLysozymeContact lensFluorescenceChemistryLens (geology)BiophysicsChromatographyBiochemistryOpticsBiologyPhysics

Abstract

fetched live from OpenAlex

PURPOSE: To optimize a fluorescence-based lysozyme activity assay to investigate the conformational state of lysozyme in solution and to determine the impact of extraction and evaporation procedures and the possible interference of contact lens materials on lysozyme activity. METHODS: The fluorescence-based lysozyme activity assay, Enzchek (Molecular Probes Inc, Eugene, OR) which utilizes fluorescently quenched Micrococcus lysodeikticus, was compared to the gold standard, classical lysozyme turbidity assay, using four differently concentrated lysozyme samples (20, 10, 5.0 and 2.0 ng/µL). Furthermore, six differently concentrated lysozyme samples (2.0, 1.0, 0.5, 0.25, 0.125 and 0.01 µg/µL) were quantified using the fluorescence-based assay in the presence of extraction solvents consisting of 0.2% and 0.02% trifluroacetic acid/acetonitrile and following evaporation procedures. RESULTS: A standard curve was generated by the fluorescence-based assay ranging from 2 to 150 ng. The total active lysozyme quantified in the four lysozyme samples was not significantly different between the two assays (p > 0.05) and the concordance correlation coefficient was determined to be 0.995. However an average discrepancy between the two assays was found to be 0.474 ng, with the turbidity assay typically reporting higher active lysozyme measurements. The sensitivity of the fluorescence-based assay was higher than the classical turbidity assay when quantifying 20 ng or less active lysozyme. Following the extraction and evaporation procedures and the addition of lens extracts, the total active lysozyme recovered was 95% or greater. CONCLUSIONS: In comparison to the classical turbidity assay, the fluorescence-based assay is a very sensitive method, making it a favorable technique, particularly when studying contact lens materials that deposit relatively low levels of lysozyme.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.003
metaresearch head score (Gemma)0.003
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Methods · Consensus signal: none
Teacher disagreement score0.003
Threshold uncertainty score0.016

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0030.003
Meta-epidemiology (narrow)0.0020.000
Meta-epidemiology (broad)0.0010.001
Bibliometrics0.0010.001
Science and technology studies0.0000.001
Scholarly communication0.0010.001
Open science0.0010.001
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0010.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.193
GPT teacher head0.450
Teacher spread0.257 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreMethods

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations18
Published2013
Admission routes2
Has abstractyes

Explore more

Same venueCurrent Eye ResearchSame topicOcular Surface and Contact LensFrench-language works237,207