MétaCan
Menu
Back to cohort
Record W2102600146 · doi:10.4037/ccn2009607

Electrocardiography Pitfalls and Artifacts: The 10 Commandments

2009· article· en· W2102600146 on OpenAlex

Why this work is in the frame

A frame that forgets how it found something cannot be audited. These are the routes that admitted this work.

affAt least one author lists a Canadian institution in the pinned OpenAlex snapshot.

Bibliographic record

VenueCritical Care Nurse · 2009
Typearticle
Languageen
FieldMedicine
TopicHealthcare Technology and Patient Monitoring
Canadian institutionsQueen's University
Fundersnot available
KeywordsMedicineElectrocardiographyTen CommandmentsCardiology

Abstract

fetched live from OpenAlex

What potential pitfalls can adversely affect the interpretation of 12-lead ECGs?Many potential pitfalls can adversely affect the interpretation of 12-lead ambulatory and telemetry electrocardiograms (ECGs). Artifacts, for example, are a common finding in patients who require ECG monitoring. Artifacts are defined as ECG abnormalities that may be due to sources other than the electrical activity of the heart. Failure to correctly distinguish between an arrhythmia and artifact can result in misdiagnosis and unnecessary therapeutic interventions.1The most common causes of artifacts originate from internal (physiological) and external (non-physiological) sources (Table 1). Artifacts created from these sources can simulate arrhythmias such as atrial flutter and ventricular tachycardia.2Electrode misplacement is another common artifact. Such misplacement may lead to changes in ECG morphology that could potentially be interpreted as ischemic in origin.3 Electrode misplacements can also mimic serious arrhythmias and lead to misdirected therapeutic decisions.4 Electrode misplacement is a relatively frequent finding in ECGs done in outpatient clinics (0.4%) and is even more common in intensive care units (4%).4 The correct position for precordial ECG electrodes is illustrated in Figure 1.Several telltale clues can help clinicians identify potential signs of electrode misplacements and artifacts. In this article, we introduce an algorithm that we developed to assist nurses and physicians in rapidly recognizing those clues and review 10 of the most common ECG pitfalls and artifacts.The indicators of electrode misplacements or artifacts that clinicians need to look for can be easily remembered by using the mnemonic REVERSE (Table 2). With this mnemonic in mind, careful and systematic examination of ECGs will help rule out problems with the recording.We describe the 10 most common ECG pitfalls and artifacts seen in our practice, presented in the form of commandments. All the ECG examples provided for this review were run at 25 mm/s, 10 mm/mV, and 100 Hz.Reversing the electrodes is one of the most common errors made when placing the ECG on a patient. Such reversal produces leads I and AVL with reverse polarity of all normal deflections (negative P wave, QRS complex, and T wave). In addition, polarity is reversed in lead AVR (positive P and QRS; Figure 2). The differential diagnosis is dextrocardia (the heart is positioned on the right side). In dextrocardia, however, the progression of the R wave in pre-cordial leads is reversed, whereas with electrode reversal, the progression is normal.5The possibility of tremor or other interference inducing an artifact that mimics ventricular tachycardia should be considered when the ECG does not match the patient’s clinical findings. A normal heart rate obtained by pulse or auscultation in an asymptomatic patient at the same time the ECG shows apparent ventricular tachycardia confirms the diagnosis. Reduction of the tremor by holding the limb or placing the electrodes on the torso will reduce interference.2,6,7 “Tracking” the R-R intervals is helpful if they can be identified before the pseudo– ventricular tachycardia. Look for R-R intervals that continue into the wide complex rhythm to see the presence of normal ventricular depolarizations throughout the pseudo–ventricular tachycardia. With careful measuring to see where normal beats should be, they will often “jump out” at the observer and become obvious, whereas at first glance they may be completely obscured. Pseudo–ventricular tachycardia has 3 characteristic signs8 (Figure 3):Amplitude of the P wave in lead I greater than in lead II and/or P-wave terminal positive component in lead III (Abdollah sign) will confirm reversal of the left arm and left leg leads.9 Confirmation with a second ECG is usually required (Figure 4).The most common reversal of the precordial leads is an exchange of V1 and V6. The way to recognize this problem is by assessing the R-wave progression in the precordial leads. Normally, the R wave will increase its amplitude from V1 to V6 and the S wave will decrease its amplitude. In the reversal situation, a tall R wave can be seen in V1 and a deep S wave in V610–13 Potential diagnostic misinterpretations include right bundle branch block, old posterior myocardial infarction, right ventricular hypertrophy, and left-sided accessory pathways (Figure 5).Tremor-induced artifact may mimic supraventricular arrhythmias (atrial flutter/atrial fibrillation) or if the artifact has sufficient amplitude, it can also mimic ventricular tachycardia and ventricular fibrillation. The correct diagnosis can be made on the basis of simple observations such as the presence of the pseudoarrhythmia when the patient moves (tremor). Careful analysis may reveal discrete components of the QRS complexes (matching the previous R-R intervals if present) through the pseudoarrhythmia (“notches sign”).8,14–16 Misinterpretation of tremor-induced artifact may lead to serious medical errors such as the initiation of long-term use of anticoagulants for pseudo– atrial fibrillation.17 Figure 6 shows pseudo–atrial flutter that was a tremor-induced artifact.Electromagnetic interference (EMI) with medical devices by cell phones is a well-recognized problem.18–21 Even though considerable controversy remains about the use of cell phones in hospitals, the evidence is clear that cell phones can produce EMI with many different medical devices (eg, ECG monitor, ventilator, infusion pump, dialysis machine, apnea monitor, external pacemaker, internal pacemaker, and defibrillator). The ability of a cell phone or a wireless device to induce EMI depends on the distance, the ability of medical equipment to resist EMI, and the technology of the cell phone (digital vs analog, which are the 2 basic systems cell phones use to operate, and single-band of operation vs dual and frequency band of operation).19 As shown by previous investigators,19 a 1-m (3.28 ft) distance between the source of EMI and medical devices safely eliminates EMI. Only a few cases in which cell phones and wireless devices interfered with ECG machines have been reported.20 We simulated a case created in our laboratory by activating a cell phone (digital) less than 25 cm (9.8 in) from the ECG machine acquisition module (MAC 5000 Resting ECG Analysis System, GE Medical Systems, Waukesha, Wisconsin). The rapid, sharp, and low-amplitude signals disappeared when the cell phone was removed or deactivated (Figure 7). ECG technicians and nurses should avoid using cell phones when they are recording ECGs until further research in this area is available. This limitation may also have implications for paramedics and ambulance attendants who obtain and interpret ECGs on patients in the field.Electrodes are placed on the torso near the extremities rather than on the limbs for different reasons. During an emergency, placing leads on the torso reduces the time needed for undressing the patient and in most cases will allow a correct ECG diagnosis.22 However, in most circumstances, the torso position should not replace the standard position on the limbs. The torso position induces a change in how the electrical vectors are recorded. Pseudo–Q waves and pseudo–ST-segment elevation in the inferior leads5,23,24 could potentially be misinterpreted as myocardial infarction (Figure 8).Placing the telemetry electrodes on top of the ECG electrodes or vice versa is a common mistake. Usually, telemetry electrodes are placed in the same region where the ECG electrodes need to be placed. This superimposition of electrodes may create a distortion of the ST segment that mimics ST-segment elevation or arrhythmias due to EMI of the telemetry on the ECG machine25 (Figure 9).If a reversal involves the right leg and one of the arms, the recording will be zero potential difference between the legs.5 This pseudoasystole in an isolated lead may occur in lead II (reversal of right arm and right leg electrodes) or in lead III (reversal of left arm and right leg electrodes; Figure 10).In some clinical situations (eg, decompensated heart failure, respiratory insufficiency, orthopedic limitations), the ECG must be recorded with the patient sitting upright or in a semi-Fowler’s position. Changing the body position can affect the QRS axis and QRS amplitude.26,27 Currently, no distinctions in the recording methods need to be made when an ECG is recorded with the patient sitting (~90°); however, recognizing slight alterations of the QRS complex may be helpful to avoid wrong interpretations. An annotation indicating the position of the patient (if different than usual) may be helpful for the physician interpreting the recordings. In this case, note the reduction of the QRS amplitude in lead III, which is a lead that is particularly sensitive to changes in diaphragmatic position (Figure 11).The ECG is one of the most valuable tools in our daily practice. Many health care providers interpret ECGs and initiate therapeutic interventions on the basis of such interpretations. Recognizing ECG artifacts and other pitfalls will enable clinicians to avoid unnecessary therapeutic interventions and may allow them to correct the recording methods to obtain a proper ECG.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.559
Threshold uncertainty score0.283

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.021
GPT teacher head0.358
Teacher spread0.337 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it