MétaCan
Menu
Retour à la cohorte
Enregistrement W2102600146 · doi:10.4037/ccn2009607

Electrocardiography Pitfalls and Artifacts: The 10 Commandments

2009· article· en· W2102600146 sur OpenAlex

Pourquoi ce travail est dans la base

Une base qui oublie comment elle a trouvé un travail ne peut pas être vérifiée. Voici les voies qui ont admis celui-ci.

affAu moins un auteur déclare une institution canadienne dans l'instantané OpenAlex épinglé.

Notice bibliographique

RevueCritical Care Nurse · 2009
Typearticle
Langueen
DomaineMedicine
ThématiqueHealthcare Technology and Patient Monitoring
Établissements canadiensQueen's University
Organismes subventionnairesnon disponible
Mots-clésMedicineElectrocardiographyTen CommandmentsCardiology

Résumé

récupéré en direct d'OpenAlex

What potential pitfalls can adversely affect the interpretation of 12-lead ECGs?Many potential pitfalls can adversely affect the interpretation of 12-lead ambulatory and telemetry electrocardiograms (ECGs). Artifacts, for example, are a common finding in patients who require ECG monitoring. Artifacts are defined as ECG abnormalities that may be due to sources other than the electrical activity of the heart. Failure to correctly distinguish between an arrhythmia and artifact can result in misdiagnosis and unnecessary therapeutic interventions.1The most common causes of artifacts originate from internal (physiological) and external (non-physiological) sources (Table 1). Artifacts created from these sources can simulate arrhythmias such as atrial flutter and ventricular tachycardia.2Electrode misplacement is another common artifact. Such misplacement may lead to changes in ECG morphology that could potentially be interpreted as ischemic in origin.3 Electrode misplacements can also mimic serious arrhythmias and lead to misdirected therapeutic decisions.4 Electrode misplacement is a relatively frequent finding in ECGs done in outpatient clinics (0.4%) and is even more common in intensive care units (4%).4 The correct position for precordial ECG electrodes is illustrated in Figure 1.Several telltale clues can help clinicians identify potential signs of electrode misplacements and artifacts. In this article, we introduce an algorithm that we developed to assist nurses and physicians in rapidly recognizing those clues and review 10 of the most common ECG pitfalls and artifacts.The indicators of electrode misplacements or artifacts that clinicians need to look for can be easily remembered by using the mnemonic REVERSE (Table 2). With this mnemonic in mind, careful and systematic examination of ECGs will help rule out problems with the recording.We describe the 10 most common ECG pitfalls and artifacts seen in our practice, presented in the form of commandments. All the ECG examples provided for this review were run at 25 mm/s, 10 mm/mV, and 100 Hz.Reversing the electrodes is one of the most common errors made when placing the ECG on a patient. Such reversal produces leads I and AVL with reverse polarity of all normal deflections (negative P wave, QRS complex, and T wave). In addition, polarity is reversed in lead AVR (positive P and QRS; Figure 2). The differential diagnosis is dextrocardia (the heart is positioned on the right side). In dextrocardia, however, the progression of the R wave in pre-cordial leads is reversed, whereas with electrode reversal, the progression is normal.5The possibility of tremor or other interference inducing an artifact that mimics ventricular tachycardia should be considered when the ECG does not match the patient’s clinical findings. A normal heart rate obtained by pulse or auscultation in an asymptomatic patient at the same time the ECG shows apparent ventricular tachycardia confirms the diagnosis. Reduction of the tremor by holding the limb or placing the electrodes on the torso will reduce interference.2,6,7 “Tracking” the R-R intervals is helpful if they can be identified before the pseudo– ventricular tachycardia. Look for R-R intervals that continue into the wide complex rhythm to see the presence of normal ventricular depolarizations throughout the pseudo–ventricular tachycardia. With careful measuring to see where normal beats should be, they will often “jump out” at the observer and become obvious, whereas at first glance they may be completely obscured. Pseudo–ventricular tachycardia has 3 characteristic signs8 (Figure 3):Amplitude of the P wave in lead I greater than in lead II and/or P-wave terminal positive component in lead III (Abdollah sign) will confirm reversal of the left arm and left leg leads.9 Confirmation with a second ECG is usually required (Figure 4).The most common reversal of the precordial leads is an exchange of V1 and V6. The way to recognize this problem is by assessing the R-wave progression in the precordial leads. Normally, the R wave will increase its amplitude from V1 to V6 and the S wave will decrease its amplitude. In the reversal situation, a tall R wave can be seen in V1 and a deep S wave in V610–13 Potential diagnostic misinterpretations include right bundle branch block, old posterior myocardial infarction, right ventricular hypertrophy, and left-sided accessory pathways (Figure 5).Tremor-induced artifact may mimic supraventricular arrhythmias (atrial flutter/atrial fibrillation) or if the artifact has sufficient amplitude, it can also mimic ventricular tachycardia and ventricular fibrillation. The correct diagnosis can be made on the basis of simple observations such as the presence of the pseudoarrhythmia when the patient moves (tremor). Careful analysis may reveal discrete components of the QRS complexes (matching the previous R-R intervals if present) through the pseudoarrhythmia (“notches sign”).8,14–16 Misinterpretation of tremor-induced artifact may lead to serious medical errors such as the initiation of long-term use of anticoagulants for pseudo– atrial fibrillation.17 Figure 6 shows pseudo–atrial flutter that was a tremor-induced artifact.Electromagnetic interference (EMI) with medical devices by cell phones is a well-recognized problem.18–21 Even though considerable controversy remains about the use of cell phones in hospitals, the evidence is clear that cell phones can produce EMI with many different medical devices (eg, ECG monitor, ventilator, infusion pump, dialysis machine, apnea monitor, external pacemaker, internal pacemaker, and defibrillator). The ability of a cell phone or a wireless device to induce EMI depends on the distance, the ability of medical equipment to resist EMI, and the technology of the cell phone (digital vs analog, which are the 2 basic systems cell phones use to operate, and single-band of operation vs dual and frequency band of operation).19 As shown by previous investigators,19 a 1-m (3.28 ft) distance between the source of EMI and medical devices safely eliminates EMI. Only a few cases in which cell phones and wireless devices interfered with ECG machines have been reported.20 We simulated a case created in our laboratory by activating a cell phone (digital) less than 25 cm (9.8 in) from the ECG machine acquisition module (MAC 5000 Resting ECG Analysis System, GE Medical Systems, Waukesha, Wisconsin). The rapid, sharp, and low-amplitude signals disappeared when the cell phone was removed or deactivated (Figure 7). ECG technicians and nurses should avoid using cell phones when they are recording ECGs until further research in this area is available. This limitation may also have implications for paramedics and ambulance attendants who obtain and interpret ECGs on patients in the field.Electrodes are placed on the torso near the extremities rather than on the limbs for different reasons. During an emergency, placing leads on the torso reduces the time needed for undressing the patient and in most cases will allow a correct ECG diagnosis.22 However, in most circumstances, the torso position should not replace the standard position on the limbs. The torso position induces a change in how the electrical vectors are recorded. Pseudo–Q waves and pseudo–ST-segment elevation in the inferior leads5,23,24 could potentially be misinterpreted as myocardial infarction (Figure 8).Placing the telemetry electrodes on top of the ECG electrodes or vice versa is a common mistake. Usually, telemetry electrodes are placed in the same region where the ECG electrodes need to be placed. This superimposition of electrodes may create a distortion of the ST segment that mimics ST-segment elevation or arrhythmias due to EMI of the telemetry on the ECG machine25 (Figure 9).If a reversal involves the right leg and one of the arms, the recording will be zero potential difference between the legs.5 This pseudoasystole in an isolated lead may occur in lead II (reversal of right arm and right leg electrodes) or in lead III (reversal of left arm and right leg electrodes; Figure 10).In some clinical situations (eg, decompensated heart failure, respiratory insufficiency, orthopedic limitations), the ECG must be recorded with the patient sitting upright or in a semi-Fowler’s position. Changing the body position can affect the QRS axis and QRS amplitude.26,27 Currently, no distinctions in the recording methods need to be made when an ECG is recorded with the patient sitting (~90°); however, recognizing slight alterations of the QRS complex may be helpful to avoid wrong interpretations. An annotation indicating the position of the patient (if different than usual) may be helpful for the physician interpreting the recordings. In this case, note the reduction of the QRS amplitude in lead III, which is a lead that is particularly sensitive to changes in diaphragmatic position (Figure 11).The ECG is one of the most valuable tools in our daily practice. Many health care providers interpret ECGs and initiate therapeutic interventions on the basis of such interpretations. Recognizing ECG artifacts and other pitfalls will enable clinicians to avoid unnecessary therapeutic interventions and may allow them to correct the recording methods to obtain a proper ECG.

Récupéré en direct depuis OpenAlex et désinversé. Les résumés ne sont pas conservés dans cette base de données : les index inversés représentent 8,6 Go des 9,3 Go de texte de la base, et le serveur dispose de 13 Go libres.

Prédiction distillée sur la base complète

Imitation des enseignants

Ni prévalence calibrée, ni vérité terrain. Validation humaine à venir. Apprise à partir de 10 348 étiquettes directes de Codex et de 10 348 étiquettes directes de Gemma. Le mode candidate est l'union des têtes enseignantes seuillées; le consensus est leur intersection. Ces sorties portent le statut machine_predicted_unvalidated et ne sont ni des étiquettes humaines ni des étiquettes directes de modèles de pointe.

score de la tête « metaresearch » (Codex)0,000
score de la tête « metaresearch » (Gemma)0,000
Version: codex-gemma-dda1882f352aStatut de validation: machine_predicted_unvalidated
Catégories candidatesaucune
Catégories consensuellesaucune
DomaineSignal candidat: aucune · Signal consensuel: aucune
Devis d'étudeSignal candidat: Observationnel · Signal consensuel: aucune
GenreSignal candidat: Empirique · Signal consensuel: Empirique
Score de désaccord entre enseignants0,559
Score d'incertitude au seuil0,283

Scores Codex et Gemma par catégorie

CatégorieCodexGemma
Métarecherche0,0000,000
Méta-épidémiologie (sens strict)0,0000,000
Méta-épidémiologie (sens large)0,0000,000
Bibliométrie0,0000,000
Études des sciences et des technologies0,0000,000
Communication savante0,0000,000
Science ouverte0,0000,000
Intégrité de la recherche0,0000,000
Charge utile insuffisante (le modèle a refusé de juger)0,0000,000

Scores machine (provisoires)

Les deux têtes enseignantes du modèle étudiant, lues sur ce travail. Un score ordonne la base pour la relecture; il n'affirme jamais une catégorie, et le statut de validation accompagne chaque rangée tel quel.

Scores de référence d'un modèle non mature (critères de maturité non atteints, 7 itérations). Un score ordonne; il n'affirme jamais une catégorie.

Tête enseignante Opus0,021
Tête enseignante GPT0,358
Écart entre enseignants0,337 · la distance entre les deux têtes enseignantes sur ce seul travail
Statut de validationscore_only:v0-immature-baseline · tel quel depuis la passe de notation : score_only signifie que le nombre peut ordonner les travaux, et qu'aucune étiquette de catégorie n'en découle