MétaCan
Menu
Back to cohort
Record W2336292745 · doi:10.1007/s40800-016-0026-8

Difficulty Using Smart Pump Logs to Recreate a Patient Safety Event: Case Study and Considerations for Pump Enhancements

2016· article· en· W2336292745 on OpenAlexaff
Andrew Ibey, Derek Sayre Andrews, Barb Ferreira

Bibliographic record

VenueDrug Safety - Case Reports · 2016
Typearticle
Languageen
FieldMedicine
TopicHealthcare Technology and Patient Monitoring
Canadian institutionsLions Gate HospitalVancouver Coastal HealthVancouver General HospitalProvidence Health Care
Fundersnot available
KeywordsTroubleshootingMedicineInfusion pumpCardiopulmonary resuscitationMedical emergencyPatient safetyCoronary care unitMyocardial infarctionAnesthesiaComputer scienceResuscitationCardiologyHealth care

Abstract

fetched live from OpenAlex

The authors present a case in which a physical anomaly with an infusion pump resulted in an unforeseen fault that the nurse's attempts to resolve unknowingly exacerbated. This case study presents the first report in the literature to detail the difficulty in recreating a patient safety event using smart pump logs, support server continuous quality improvement (CQI) data, and the drug order entry system to elucidate the clinical scenario. A 75-year-old male patient presented to a major teaching hospital and was admitted to the intensive care unit (ICU) with a massive gastrointestinal bleed and myocardial infarction, then stabilized. One of the patient's pumps alarmed "communication error" on the display. The display gave no explicit instructions about how to resolve the issue, and resolution was not intuitive. Attempts to clear the alarm failed, so the module was disconnected to reprogram the infusion, causing an interruption in the dopamine. Over the course of approximately 2 min of troubleshooting, the patient's blood pressure decreased from 109/50 to 60/30, with a rapid pulse change from a consistent 95 up to 115 and subsequently 135 beats per minute. A cardiac arrest ensued and a code blue was called. All cardiac drugs, including the dopamine, were suspended during the code. Cardiopulmonary resuscitation was performed and the patient survived the code. Post-code, the dopamine and epinephrine were restarted, and the norepinephrine was discontinued. The patient's condition remained very unstable. Pump logs and the server database were queried to locate relevant equipment. It was concluded that dirty contacts on the inter-unit interface (IUI) connectors between the PC unit (PCU) and the modules caused the alarm message "communication error" to appear on the PCU display. Learning yielded a nursing practice alert to clarify how a nurse should resolve a "communication error", and appropriate cleaning protocols were promptly implemented. The investigation found smart pump event logs and proprietary software are not designed with any forethought as to retrospective reconstruction of incident investigations, leaving facilities to cobble together pieces of information from multiple sources to determine what occurred. The authors also suggest further pump enhancements, challenging pump manufacturers to go to the next level of integration and enable greater patient safety with smart infusion pumps.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.001
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Case report · Consensus signal: Case report
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.335
Threshold uncertainty score0.860

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0010.001
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0000.000
Science and technology studies0.0010.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.054
GPT teacher head0.349
Teacher spread0.294 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designCase report
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations3
Published2016
Admission routes1
Has abstractyes

Explore more

Same venueDrug Safety - Case ReportsSame topicHealthcare Technology and Patient MonitoringFrench-language works237,207