Five-Year Safety and Performance Results from the Argus II Retinal Prosthesis System Clinical Trial
Why is this work in the frame?
A frame that forgets how it found something cannot be audited. These are the routes that admitted this work.
No Canadian affiliation. An affiliation-only frame — the usual design — would never have seen this work. It is one of the works that make the case for inverting the frame.
Machine scores (provisional)
Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.
The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.
- Teacher spread
- 0.240 · how far apart the two teachers sit on this one work
- Validation status
score_only:v0-immature-baseline· verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it
Abstract
PURPOSE: The Argus II Retinal Prosthesis System (Second Sight Medical Products, Inc, Sylmar, CA) was developed to restore some vision to patients blind as a result of retinitis pigmentosa (RP) or outer retinal degeneration. A clinical trial was initiated in 2006 to study the long-term safety and efficacy of the Argus II System in patients with bare or no light perception resulting from end-stage RP. DESIGN: Prospective, multicenter, single-arm clinical trial. Within-patient controls included the nonimplanted fellow eye and patients' native residual vision compared with their vision with the Argus II. PARTICIPANTS: Thirty participants in 10 centers in the United States and Europe. METHODS: The worse-seeing eye of blind patients was implanted with the Argus II. Patients wore glasses mounted with a small camera and a video processor that converted images into stimulation patterns sent to the electrode array on the retina. MAIN OUTCOME MEASURES: The primary outcome measures were safety (the number, seriousness, and relatedness of adverse events) and visual function, as measured by 3 computer-based, objective tests. Secondary measures included functional vision performance on objectively scored real-world tasks. RESULTS: Twenty-four of 30 patients remained implanted with functioning Argus II Systems at 5 years after implantation. Only 1 additional serious adverse event was experienced after the 3-year time point. Patients performed significantly better with the Argus II on than off on all visual function tests and functional vision tasks. CONCLUSIONS: The 5-year results of the Argus II trial support the long-term safety profile and benefit of the Argus II System for patients blind as a result of RP. The Argus II is the first and only retinal implant to have market approval in the European Economic Area, the United States, and Canada.
Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.
The record
- Venue
- Ophthalmology
- Topic
- Neuroscience and Neural Engineering
- Field
- Neuroscience
- Canadian institutions
- —
- Funders
- National Eye Institute
- Keywords
- Visual prosthesisArgusRetinal implantMedicineRetinitis pigmentosaAdverse effectClinical trialVisual acuitySurgeryOphthalmologyImplantRetinalInternal medicine
- Has abstract in OpenAlex
- yes