A comparison of pulse oximetry and laser Doppler flowmetry in monitoring sequential vascular occlusion in a rabbit ear model

2003· article· en· W254406604 on OpenAlexvenueno aff
Geoffrey G. Hallock, David C. Rice

Bibliographic record

VenueCanadian Journal of Plastic Surgery · 2003
Typearticle
Languageen
FieldMedicine
TopicReconstructive Surgery and Microvascular Techniques
Canadian institutionsnot available
Fundersnot available
KeywordsMedicinePulse oximetryLaser Doppler velocimetryOcclusionPerioperativeVascular occlusionAnastomosisAnesthesiaSurgeryCardiologyBlood flow

Abstract

fetched live from OpenAlex

The ultimate success of any free flap transfer depends not only on the experience of the given surgical team, but also requires constant vigilance in the perioperative period to assure continued anastomotic patency. Clinical acumen remains the 'industry' standard, but adjunctive monitoring devices are important to reinforce these subjective evaluations. We routinely use laser Doppler flowmetry in spite of known drawbacks, including expense. On the other hand, we know that pulse oximetry is a possible and readily available alternative. The efficacy of these two monitoring systems was directly compared in a New Zealand white rabbit central artery and vein ear model. Arterial occlusion was immediately recognized by a precipitous drop in flow and oxygen saturation (SaO2). A rapid drop in flow after venous occlusion with a trend to zero was also noted with laser Doppler flowmetry. The SaO2 had a very slow but steady gradual decline, but values remained in the range of normoxemia for a prolonged time. Because a minimum threshold of SaO2 for predicting venous occlusion was elusive, the usefulness of pulse oximetry for monitoring microsurgical composite tissue transfers is limited.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.002
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: Observational
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.142
Threshold uncertainty score0.621

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0010.002
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0020.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.033
GPT teacher head0.285
Teacher spread0.251 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations8
Published2003
Admission routes1
Has abstractyes

Explore more

Same venueCanadian Journal of Plastic SurgerySame topicReconstructive Surgery and Microvascular TechniquesFrench-language works237,207