MétaCan
Menu
Back to cohort
Record W2583762396 · doi:10.1016/j.hrcr.2015.04.007

Inappropriate shocks by subcutaneous implantable cardioverter-defibrillator due to T-wave oversensing in hyperkalemia leading to ventricular fibrillation

2015· article· en· W2583762396 on OpenAlexaff
Omid Kiamanesh, Deirdre O’Neill, Soori Sivakumaran, Shane Kimber

Bibliographic record

VenueHeartRhythm Case Reports · 2015
Typearticle
Languageen
FieldMedicine
TopicCardiac pacing and defibrillation studies
Canadian institutionsUniversity of Alberta
Fundersnot available
KeywordsMedicineCardiologyHyperkalemiaInternal medicineVentricular fibrillationImplantable cardioverter-defibrillator

Abstract

fetched live from OpenAlex

IntroductionTabled 1KEY TEACHING POINTS•The most frequent complication of the subcutaneous implantable cardioverter-defibrillator (S-ICD) is inappropriate shocks. In contrast to transvenous ICDs, inappropriate shocks from S-ICDs are often due to T-wave oversensing.•T-wave oversensing by the S-ICD may occur in the context of hyperkalemia.•The population in which there may be a preference for an S-ICD may also be at the greatest risk of the limitations of rhythm detection and discrimination. Open table in a new tab The implantable cardioverter-defibrillator (ICD) has reduced mortality in survivors of sudden cardiac arrest and patients at high risk of sudden cardiac death.1Ezekowitz J.A. Armstrong P.W. McAlister F.A. Implantable cardioverter defibrillators in primary and secondary prevention: a systematic review of randomized, controlled trials.Ann Intern Med. 2003; 138: 445-452Crossref PubMed Scopus (219) Google Scholar This benefit comes at the cost of device-related complications, including those related to transvenous leads.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar The subcutaneous ICD (S-ICD) is an alternative to the conventional transvenous ICD (TV-ICD) system and has no transvenous leads, thereby avoiding endocardial lead–related complications.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar However, with only subcutaneous electrodes, rhythm detection and discrimination is a much more challenging task for the S-ICD. We describe a case of ventricular fibrillation induced by inappropriate shocks from the S-ICD owing to T-wave oversensing (TWOS) in the setting of hyperkalemia and dialysis-dependent chronic renal failure.Case reportA 33-year-old man with a history of dilated cardiomyopathy with reduced left ventricular ejection fraction (25%) and end-stage renal disease requiring intermittent hemodialysis was resuscitated in the community after experiencing rapid monomorphic ventricular tachycardia with syncope. He received a single-chamber TV-ICD (Protecta XT VR; Medtronic, Minneapolis, MN) for secondary prevention prior to hospital discharge. During the patient’s routine assessments in the device clinic, intermittent TWOS was noted, but in each instance it had been appropriately discriminated by the TWOS algorithms of the device and no inappropriate therapies had been delivered. In follow-up, the patient developed a persistent coagulase-negative Staphylococcus bacteremia, which ultimately required the removal of the ICD system 13 months after the original implant.The patient required a chronic indwelling catheter for hemodialysis; therefore, to mitigate risk of future infection following resolution of the patient’s bacteremia, the patient received an S-ICD device implantation (SQ-RX Pulse Generator Model #1010 and Q-TRAK Subcutaneous Electrode Model #3010; Cameron Health/Boston Scientific, San Clemente, CA). The generator was placed at the cardiac apex/anterior axillary line and the subcutaneous electrode along the right sternal border. There was good QRS-T complex discrimination at implantation (Figure 1) and defibrillation threshold testing was successful with reverse polarity at 80 joules.Approximately 1 month after S-ICD placement, the patient presented to hospital with complaints of multiple, unprovoked device shocks. He reported missing 2 consecutive dialysis treatments and his serum potassium was found to be elevated at 7.0 mmol/L. Interrogation of the S-ICD revealed 5 episodes of tachycardia and 17 shocks delivered. Rhythm analysis showed that each reported episode of tachycardia was due to TWOS. During the first 3 episodes, between 3 and 5 inappropriate shocks were delivered during sinus rhythm. During the fourth and fifth episodes, the inappropriate shocks were delivered on the T-wave and induced nonsustained polymorphic ventricular tachycardia and ventricular fibrillation, respectively (Figure 2). The ventricular fibrillation lasted approximately 87 seconds and required 4 shocks before the patient returned to normal sinus rhythm. Urgent hemodialysis was initiated for correction of hyperkalemia and no further shocks occurred. The conditional shock zone and shock zone were increased to 200 and 230 beats per minute, respectively. Compliance with dialysis was recommended and the patient was discharged with clinical follow-up.Figure 2Surface electrocardiogram from the subcutaneous implantable cardioverter-defibrillator demonstrating inappropriate shocks due to T-wave oversensing in the setting of hyperkalemia (serum potassium 7.0 mmol/L), leading to sustained ventricular fibrillation requiring 4 shocks prior to termination.View Large Image Figure ViewerDownload (PPT)DiscussionThe S-ICD system is a novel technology that offers a less invasive alternative to the conventional TV-ICD system.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar Preference may be given to the S-ICD in patients for whom there is a desire to avoid the complications associated with intracardiac leads. These patients may include the young, those with poor vascular access, and those at high risk of bacteremia, such as patients with chronic indwelling endovascular catheters or those receiving hemodialysis. However, long-term safety data for the S-ICD are lacking in these complex patients.The most frequent complication of the S-ICD is inappropriate shocks.3Weiss R. Knight B.P. Gold M.R. Leon A.R. Herre J.M. Hood M. Rashtian M. Kremers M. Crozier I. Lee K.L. Smith W. Burke M.C. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator.Circulation. 2013; 128: 944-953Crossref PubMed Scopus (387) Google Scholar, 4Lambiase P.D. Barr C. Theuns D.A. et al.Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry.Eur Heart J. 2014; 35: 1657-1665Crossref PubMed Scopus (350) Google Scholar While inappropriate shocks from TV-ICDs are most often due to supraventricular arrhythmias, inappropriate shocks from S-ICDs are most often due to oversensing of T-waves and low-amplitude cardiac signals.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google ScholarTransient hyperkalemia is common in dialysis patients, a population that is difficult to manage and in which an S-ICD might be more likely to be considered owing to a decreased risk of infection. Unfortunately, hyperkalemia represents one mechanism that may lead to TWOS in the S-ICD. Increased extracellular potassium concentrations may decrease the amplitude of the QRS complex and increase the amplitude of the T-wave. Hyperkalemia may also increase defibrillation thresholds.5Sims J.J. Miller A.W. Ujhelyi M.R. Regional hyperkalemia increases ventricular defibrillation energy requirements.J Cardiovasc Electrophysiol. 2000; 11: 634-641Crossref PubMed Scopus (28) Google Scholar In this case, the impaired ability of the S-ICD to discriminate the QRS-T complex led to the induction of ventricular fibrillation after a shock was delivered on the T-wave. The patient required 4 shocks at maximum device outputs to terminate ventricular fibrillation.ConclusionThe population in which there may be a preference for an S-ICD may also be the population most at risk of the limitations of this technology. Further enhancement of the rhythm detection/discrimination abilities of these devices will significantly enhance their clinical utility. Higher defibrillation safety margins at implantation may also be warranted. To our knowledge, this is the first report of TWOS in the setting of hyperkalemia leading to inappropriate shocks from an S-ICD, resulting in sustained ventricular fibrillation. IntroductionTabled 1KEY TEACHING POINTS•The most frequent complication of the subcutaneous implantable cardioverter-defibrillator (S-ICD) is inappropriate shocks. In contrast to transvenous ICDs, inappropriate shocks from S-ICDs are often due to T-wave oversensing.•T-wave oversensing by the S-ICD may occur in the context of hyperkalemia.•The population in which there may be a preference for an S-ICD may also be at the greatest risk of the limitations of rhythm detection and discrimination. Open table in a new tab The implantable cardioverter-defibrillator (ICD) has reduced mortality in survivors of sudden cardiac arrest and patients at high risk of sudden cardiac death.1Ezekowitz J.A. Armstrong P.W. McAlister F.A. Implantable cardioverter defibrillators in primary and secondary prevention: a systematic review of randomized, controlled trials.Ann Intern Med. 2003; 138: 445-452Crossref PubMed Scopus (219) Google Scholar This benefit comes at the cost of device-related complications, including those related to transvenous leads.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar The subcutaneous ICD (S-ICD) is an alternative to the conventional transvenous ICD (TV-ICD) system and has no transvenous leads, thereby avoiding endocardial lead–related complications.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar However, with only subcutaneous electrodes, rhythm detection and discrimination is a much more challenging task for the S-ICD. We describe a case of ventricular fibrillation induced by inappropriate shocks from the S-ICD owing to T-wave oversensing (TWOS) in the setting of hyperkalemia and dialysis-dependent chronic renal failure. The implantable cardioverter-defibrillator (ICD) has reduced mortality in survivors of sudden cardiac arrest and patients at high risk of sudden cardiac death.1Ezekowitz J.A. Armstrong P.W. McAlister F.A. Implantable cardioverter defibrillators in primary and secondary prevention: a systematic review of randomized, controlled trials.Ann Intern Med. 2003; 138: 445-452Crossref PubMed Scopus (219) Google Scholar This benefit comes at the cost of device-related complications, including those related to transvenous leads.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar The subcutaneous ICD (S-ICD) is an alternative to the conventional transvenous ICD (TV-ICD) system and has no transvenous leads, thereby avoiding endocardial lead–related complications.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar However, with only subcutaneous electrodes, rhythm detection and discrimination is a much more challenging task for the S-ICD. We describe a case of ventricular fibrillation induced by inappropriate shocks from the S-ICD owing to T-wave oversensing (TWOS) in the setting of hyperkalemia and dialysis-dependent chronic renal failure. Case reportA 33-year-old man with a history of dilated cardiomyopathy with reduced left ventricular ejection fraction (25%) and end-stage renal disease requiring intermittent hemodialysis was resuscitated in the community after experiencing rapid monomorphic ventricular tachycardia with syncope. He received a single-chamber TV-ICD (Protecta XT VR; Medtronic, Minneapolis, MN) for secondary prevention prior to hospital discharge. During the patient’s routine assessments in the device clinic, intermittent TWOS was noted, but in each instance it had been appropriately discriminated by the TWOS algorithms of the device and no inappropriate therapies had been delivered. In follow-up, the patient developed a persistent coagulase-negative Staphylococcus bacteremia, which ultimately required the removal of the ICD system 13 months after the original implant.The patient required a chronic indwelling catheter for hemodialysis; therefore, to mitigate risk of future infection following resolution of the patient’s bacteremia, the patient received an S-ICD device implantation (SQ-RX Pulse Generator Model #1010 and Q-TRAK Subcutaneous Electrode Model #3010; Cameron Health/Boston Scientific, San Clemente, CA). The generator was placed at the cardiac apex/anterior axillary line and the subcutaneous electrode along the right sternal border. There was good QRS-T complex discrimination at implantation (Figure 1) and defibrillation threshold testing was successful with reverse polarity at 80 joules.Approximately 1 month after S-ICD placement, the patient presented to hospital with complaints of multiple, unprovoked device shocks. He reported missing 2 consecutive dialysis treatments and his serum potassium was found to be elevated at 7.0 mmol/L. Interrogation of the S-ICD revealed 5 episodes of tachycardia and 17 shocks delivered. Rhythm analysis showed that each reported episode of tachycardia was due to TWOS. During the first 3 episodes, between 3 and 5 inappropriate shocks were delivered during sinus rhythm. During the fourth and fifth episodes, the inappropriate shocks were delivered on the T-wave and induced nonsustained polymorphic ventricular tachycardia and ventricular fibrillation, respectively (Figure 2). The ventricular fibrillation lasted approximately 87 seconds and required 4 shocks before the patient returned to normal sinus rhythm. Urgent hemodialysis was initiated for correction of hyperkalemia and no further shocks occurred. The conditional shock zone and shock zone were increased to 200 and 230 beats per minute, respectively. Compliance with dialysis was recommended and the patient was discharged with clinical follow-up. A 33-year-old man with a history of dilated cardiomyopathy with reduced left ventricular ejection fraction (25%) and end-stage renal disease requiring intermittent hemodialysis was resuscitated in the community after experiencing rapid monomorphic ventricular tachycardia with syncope. He received a single-chamber TV-ICD (Protecta XT VR; Medtronic, Minneapolis, MN) for secondary prevention prior to hospital discharge. During the patient’s routine assessments in the device clinic, intermittent TWOS was noted, but in each instance it had been appropriately discriminated by the TWOS algorithms of the device and no inappropriate therapies had been delivered. In follow-up, the patient developed a persistent coagulase-negative Staphylococcus bacteremia, which ultimately required the removal of the ICD system 13 months after the original implant. The patient required a chronic indwelling catheter for hemodialysis; therefore, to mitigate risk of future infection following resolution of the patient’s bacteremia, the patient received an S-ICD device implantation (SQ-RX Pulse Generator Model #1010 and Q-TRAK Subcutaneous Electrode Model #3010; Cameron Health/Boston Scientific, San Clemente, CA). The generator was placed at the cardiac apex/anterior axillary line and the subcutaneous electrode along the right sternal border. There was good QRS-T complex discrimination at implantation (Figure 1) and defibrillation threshold testing was successful with reverse polarity at 80 joules. Approximately 1 month after S-ICD placement, the patient presented to hospital with complaints of multiple, unprovoked device shocks. He reported missing 2 consecutive dialysis treatments and his serum potassium was found to be elevated at 7.0 mmol/L. Interrogation of the S-ICD revealed 5 episodes of tachycardia and 17 shocks delivered. Rhythm analysis showed that each reported episode of tachycardia was due to TWOS. During the first 3 episodes, between 3 and 5 inappropriate shocks were delivered during sinus rhythm. During the fourth and fifth episodes, the inappropriate shocks were delivered on the T-wave and induced nonsustained polymorphic ventricular tachycardia and ventricular fibrillation, respectively (Figure 2). The ventricular fibrillation lasted approximately 87 seconds and required 4 shocks before the patient returned to normal sinus rhythm. Urgent hemodialysis was initiated for correction of hyperkalemia and no further shocks occurred. The conditional shock zone and shock zone were increased to 200 and 230 beats per minute, respectively. Compliance with dialysis was recommended and the patient was discharged with clinical follow-up. DiscussionThe S-ICD system is a novel technology that offers a less invasive alternative to the conventional TV-ICD system.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar Preference may be given to the S-ICD in patients for whom there is a desire to avoid the complications associated with intracardiac leads. These patients may include the young, those with poor vascular access, and those at high risk of bacteremia, such as patients with chronic indwelling endovascular catheters or those receiving hemodialysis. However, long-term safety data for the S-ICD are lacking in these complex patients.The most frequent complication of the S-ICD is inappropriate shocks.3Weiss R. Knight B.P. Gold M.R. Leon A.R. Herre J.M. Hood M. Rashtian M. Kremers M. Crozier I. Lee K.L. Smith W. Burke M.C. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator.Circulation. 2013; 128: 944-953Crossref PubMed Scopus (387) Google Scholar, 4Lambiase P.D. Barr C. Theuns D.A. et al.Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry.Eur Heart J. 2014; 35: 1657-1665Crossref PubMed Scopus (350) Google Scholar While inappropriate shocks from TV-ICDs are most often due to supraventricular arrhythmias, inappropriate shocks from S-ICDs are most often due to oversensing of T-waves and low-amplitude cardiac signals.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google ScholarTransient hyperkalemia is common in dialysis patients, a population that is difficult to manage and in which an S-ICD might be more likely to be considered owing to a decreased risk of infection. Unfortunately, hyperkalemia represents one mechanism that may lead to TWOS in the S-ICD. Increased extracellular potassium concentrations may decrease the amplitude of the QRS complex and increase the amplitude of the T-wave. Hyperkalemia may also increase defibrillation thresholds.5Sims J.J. Miller A.W. Ujhelyi M.R. Regional hyperkalemia increases ventricular defibrillation energy requirements.J Cardiovasc Electrophysiol. 2000; 11: 634-641Crossref PubMed Scopus (28) Google Scholar In this case, the impaired ability of the S-ICD to discriminate the QRS-T complex led to the induction of ventricular fibrillation after a shock was delivered on the T-wave. The patient required 4 shocks at maximum device outputs to terminate ventricular fibrillation. The S-ICD system is a novel technology that offers a less invasive alternative to the conventional TV-ICD system.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar Preference may be given to the S-ICD in patients for whom there is a desire to avoid the complications associated with intracardiac leads. These patients may include the young, those with poor vascular access, and those at high risk of bacteremia, such as patients with chronic indwelling endovascular catheters or those receiving hemodialysis. However, long-term safety data for the S-ICD are lacking in these complex patients. The most frequent complication of the S-ICD is inappropriate shocks.3Weiss R. Knight B.P. Gold M.R. Leon A.R. Herre J.M. Hood M. Rashtian M. Kremers M. Crozier I. Lee K.L. Smith W. Burke M.C. Safety and efficacy of a totally subcutaneous implantable-cardioverter defibrillator.Circulation. 2013; 128: 944-953Crossref PubMed Scopus (387) Google Scholar, 4Lambiase P.D. Barr C. Theuns D.A. et al.Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD Registry.Eur Heart J. 2014; 35: 1657-1665Crossref PubMed Scopus (350) Google Scholar While inappropriate shocks from TV-ICDs are most often due to supraventricular arrhythmias, inappropriate shocks from S-ICDs are most often due to oversensing of T-waves and low-amplitude cardiac signals.2Bardy G.H. Smith W.M. Hood M.A. et al.An entirely subcutaneous implantable cardioverter–defibrillator.N Engl J Med. 2010; 363: 36-44Crossref PubMed Scopus (548) Google Scholar Transient hyperkalemia is common in dialysis patients, a population that is difficult to manage and in which an S-ICD might be more likely to be considered owing to a decreased risk of infection. Unfortunately, hyperkalemia represents one mechanism that may lead to TWOS in the S-ICD. Increased extracellular potassium concentrations may decrease the amplitude of the QRS complex and increase the amplitude of the T-wave. Hyperkalemia may also increase defibrillation thresholds.5Sims J.J. Miller A.W. Ujhelyi M.R. Regional hyperkalemia increases ventricular defibrillation energy requirements.J Cardiovasc Electrophysiol. 2000; 11: 634-641Crossref PubMed Scopus (28) Google Scholar In this case, the impaired ability of the S-ICD to discriminate the QRS-T complex led to the induction of ventricular fibrillation after a shock was delivered on the T-wave. The patient required 4 shocks at maximum device outputs to terminate ventricular fibrillation. ConclusionThe population in which there may be a preference for an S-ICD may also be the population most at risk of the limitations of this technology. Further enhancement of the rhythm detection/discrimination abilities of these devices will significantly enhance their clinical utility. Higher defibrillation safety margins at implantation may also be warranted. To our knowledge, this is the first report of TWOS in the setting of hyperkalemia leading to inappropriate shocks from an S-ICD, resulting in sustained ventricular fibrillation. The population in which there may be a preference for an S-ICD may also be the population most at risk of the limitations of this technology. Further enhancement of the rhythm detection/discrimination abilities of these devices will significantly enhance their clinical utility. Higher defibrillation safety margins at implantation may also be warranted. To our knowledge, this is the first report of TWOS in the setting of hyperkalemia leading to inappropriate shocks from an S-ICD, resulting in sustained ventricular fibrillation.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesMeta-epidemiology (narrow)
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Case report · Consensus signal: Case report
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.108
Threshold uncertainty score1.000

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0010.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0000.001
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.033
GPT teacher head0.284
Teacher spread0.252 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

Study designCase report
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations6
Published2015
Admission routes1
Has abstractyes

Explore more

Same venueHeartRhythm Case ReportsSame topicCardiac pacing and defibrillation studiesFrench-language works237,207