2D–3D registration for cranial radiation therapy using a 3D kV CBCT and a single limited field‐of‐view 2D kV radiograph

2018· article· en· W2792256465 on OpenAlexaff
Reshma Munbodh, Jonathan Knisely, David A. Jaffray, D Moseley

Bibliographic record

VenueMedical Physics · 2018
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsPrincess Margaret Cancer Centre
Fundersnot available
KeywordsImaging phantomImage registrationRadiographyArtificial intelligenceNuclear medicineComputer visionComputer scienceImage-guided radiation therapyMedical imagingComputed radiographyDigital radiographyGround truthMathematicsMedicineRadiologyImage (mathematics)Image quality

Abstract

fetched live from OpenAlex

Purpose We present and evaluate a fully automated 2D–3D intensity‐based registration framework using a single limited field‐of‐view (FOV) 2D kV radiograph and a 3D kV CBCT for 3D estimation of patient setup errors during brain radiotherapy. Methods We evaluated two similarity measures, the Pearson correlation coefficient on image intensity values (ICC) and maximum likelihood measure with Gaussian noise (MLG), derived from the statistics of transmission images. Pose determination experiments were conducted on 2D kV radiographs in the anterior–posterior (AP) and left lateral (LL) views and 3D kV CBCTs of an anthropomorphic head phantom. In order to minimize radiation exposure and exclude nonrigid structures from the registration, limited FOV 2D kV radiographs were employed. A spatial frequency band useful for the 2D–3D registration was identified from the bone‐to‐no‐bone spectral ratio (BNBSR) of digitally reconstructed radiographs (DRRs) computed from the 3D kV planning CT of the phantom. The images being registered were filtered accordingly prior to computation of the similarity measures. We evaluated the registration accuracy achievable with a single 2D kV radiograph and with the registration results from the AP and LL views combined. We also compared the performance of the 2D–3D registration solutions proposed to that of a commercial 3D–3D registration algorithm, which used the entire skull for the registration. The ground truth was determined from markers affixed to the phantom and visible in the CBCT images. Results The accuracy of the 2D–3D registration solutions, as quantified by the root mean squared value of the target registration error (TRE) calculated over a radius of 3 cm for all poses tested, was ICCAP: 0.56 mm, MLGAP: 0.74 mm, ICCLL: 0.57 mm, MLGLL: 0.54 mm, ICC (AP and LL combined): 0.19 mm, and MLG (AP and LL combined): 0.21 mm. The accuracy of the 3D–3D registration algorithm was 0.27 mm. There was no significant difference in mean TRE for the 2D–3D registration algorithms using a single 2D kV radiograph with similarity measure and image view point. There was no significant difference in mean TRE between ICCLL, MLGLL, ICC (AP and LL combined), MLG (AP and LL combined), and the 3D–3D registration algorithm despite the smaller FOV used for the 2D–3D registration. While submillimeter registration accuracy was obtained with both ICC and MLG using a single 2D kV radiograph, combining the results from the two projection views resulted in a significantly smaller (P≤0.05) mean TRE. Conclusions Our results indicate that it is possible to achieve submillimeter registration accuracy with both ICC and MLG using either single or dual limited FOV 2D kV radiographs of the head in the AP and LL views. The registration accuracy suggests that the 2D–3D registration solutions presented are suitable for the estimation of patient setup errors not only during conventional brain radiation therapy, but also during stereotactic procedures and proton radiation therapy where tighter setup margins are required.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.004
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: none
GenreCandidate signal: Methods · Consensus signal: Methods
Teacher disagreement score0.002
Threshold uncertainty score0.007

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.004
Meta-epidemiology (narrow)0.0010.001
Meta-epidemiology (broad)0.0010.001
Bibliometrics0.0010.001
Science and technology studies0.0000.000
Scholarly communication0.0010.001
Open science0.0010.002
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0020.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.023
GPT teacher head0.314
Teacher spread0.292 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreMethods

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations11
Published2018
Admission routes1
Has abstractyes

Explore more

Same venueMedical PhysicsSame topicAdvanced Radiotherapy TechniquesFrench-language works237,207