Prospective multicenter clinical trial to evaluate the safety and effectiveness of a new glistening-free one-piece acrylic toric intraocular lens

2018· article· en· W2806248332 on OpenAlexfundno aff
Mark Packer, Jon I. Williams, Gregg Feinerman, Richard S. Hope

Bibliographic record

VenueClinical ophthalmology · 2018
Typearticle
Languageen
FieldMedicine
TopicIntraocular Surgery and Lenses
Canadian institutionsnot available
FundersValeant Pharmaceuticals International
KeywordsMedicineAstigmatismOphthalmologyIntraocular lensVisual acuityRandomized controlled trialProspective cohort studyOptometrySurgeryOptics

Abstract

fetched live from OpenAlex

Purpose: Glistening formation in the intraocular lens (IOL) optic has the potential to impact quality of vision. The enVista One-Piece Hydrophobic Acrylic Spherical IOL Model MX60 (MX60 IOL) is currently the only US Food and Drug Administration-approved IOL with a label of “no glistenings”. The purpose of this prospective, multicenter, partially randomized, partially controlled, double-masked, pivotal study was to evaluate the safety and effectiveness of the enVista One-Piece Hydrophobic Acrylic MX60T Toric IOL (enVista MX60T Toric IOL). Patients and methods: Subjects (n=191) were implanted with the enVista MX60T Toric IOL (cylinder powers 1.25, 2.00, or 2.75 D) or the parent MX60 IOL (control). Eyes within the lowest range of corneal astigmatism were randomized to receive either Toric 1.25 D IOL or control IOL in a 1:1 ratio. All subjects with corneal astigmatism requiring 2.00 or 2.75 D cylinder IOLs received toric IOLs. Rotational stability, cylinder reduction, and best-corrected distance visual acuity were primary effectiveness endpoints measured at Visit 4 (120–180 days postoperatively). Results: Visit 4 mean absolute axis misalignment in the All Toric group was 4.68°±7.33°, and all subjects had ≤5° absolute rotation from Visit 3 to Visit 4. The 1.25 D group had significantly greater improvement in dioptric cylinder reduction ( P <0.001), percent cylinder reduction ( P <0.001), and mean uncorrected distance visual acuity ( P <0.001), compared to control at Visit 4. Most adverse events (AEs) were mild, with no serious AEs in the study eyes. The rates of cumulative AEs through Visit 4 were below International Organization for Standardization (ISO) standard 11979-7 AE rates. Conclusion: enVista MX60T Toric IOL is safe and effective for patients with preoperative corneal astigmatism undergoing IOL implantation. Keywords: human, IOL, astigmatism, cataract, enVista, clinical

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.003
metaresearch head score (Gemma)0.002
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Randomized trial · Consensus signal: Randomized trial
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.005
Threshold uncertainty score0.018

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0030.002
Meta-epidemiology (narrow)0.0010.001
Meta-epidemiology (broad)0.0020.001
Bibliometrics0.0000.000
Science and technology studies0.0000.001
Scholarly communication0.0010.001
Open science0.0010.000
Research integrity0.0010.002
Insufficient payload (model declined to judge)0.0050.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.124
GPT teacher head0.442
Teacher spread0.318 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designRandomized trial
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations18
Published2018
Admission routes1
Has abstractyes

Explore more

Same venueClinical ophthalmologySame topicIntraocular Surgery and LensesFrench-language works237,207