MétaCan
Menu
Back to cohort
Record W3161347321 · doi:10.1039/d0an02496a

Plasmonic and label-free real-time quantitative PCR for point-of-care diagnostics

2021· article· en· W3161347321 on OpenAlexafffund
Padideh Mohammadyousef, Miltiadis Paliouras, Mark Trifiro, Andrew G. Kirk

Bibliographic record

VenueThe Analyst · 2021
Typearticle
Languageen
FieldEngineering
TopicBiosensors and Analytical Detection
Canadian institutionsJewish General HospitalMcGill University
FundersGenome Canada
KeywordsAmpliconPoint of carePlasmonSample (material)OptoelectronicsPoint (geometry)Point-of-care testingReal-time polymerase chain reactionMaterials scienceNanotechnologyComputer sciencePolymerase chain reactionMedicineChemistryMathematicsChromatographyPathology

Abstract

fetched live from OpenAlex

DNA amplification in 20 μL total PCR sample volume. In the second part, we report an ultrasensitive real-time amplicon detection strategy which is based on cycle-by-cycle monitoring of 260 nm absorption of the PCR sample. This was accomplished by irradiating the PCR sample using a UV LED and collecting the transmitted optical power with a photodetector. The UV absorption dependency on the nucleotides' structural degree of freedom gives rise to distinctive features in the shape of UV amplification curves for the determination of PCR results, thus circumventing the need for the complicated design of target-specific probes or intercalating fluorophores. This amplicon quantification method has a high detection sensitivity of one DNA copy. This is the first demonstration of a compact plasmonic thermocycler combined with a real-time fluorophore-free quantitative amplicon detection system. The small footprint of our PCR device stems from hardware miniaturization, while abundant sample volume facilitates highly sensitive detection and fluid handling required for in-field sample analysis, thereby making it an excellent candidate for POC molecular diagnostics.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.653
Threshold uncertainty score0.219

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.013
GPT teacher head0.240
Teacher spread0.227 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations30
Published2021
Admission routes2
Has abstractyes

Explore more

Same venueThe AnalystSame topicBiosensors and Analytical DetectionFrench-language works237,207