Feasibility of fusing three‐dimensional transabdominal and transrectal ultrasound images for comprehensive intraoperative visualization of gynecologic brachytherapy applicators
Bibliographic record
Abstract
PURPOSE: In this study, we propose combining three-dimensional (3D) transrectal ultrasound (TRUS) and 3D transabdominal ultrasound (TAUS) images of gynecologic brachytherapy applicators to leverage the advantages of each imaging perspective, providing a broader field-of-view and allowing previously obscured features to be recovered. The aim of this study was to evaluate the feasibility of fusing these 3D ultrasound (US) perspectives based on the applicator geometry in a phantom prior to clinical implementation. METHODS: In proof-of-concept experiments, 3D US images of application-specific multimodality pelvic phantoms were acquired with tandem-and-ring and tandem-and-ovoids applicators using previously validated imaging systems. Two TRUS images were acquired at different insertion depths and manually fused based on the position of the ring/ovoids to broaden the TRUS field-of-view. The phantom design allowed "abdominal thickness" to be modified to represent different body habitus and TAUS images were acquired at three thicknesses for each applicator. The merged TRUS images were then combined with TAUS images by rigidly aligning applicator components and manually refining the registration using the positions of source channels and known tandem length, as well as the ring diameter for the tandem-and-ring applicator. Combined 3D US images were manually, rigidly registered to images from a second modality (magnetic resonance (MR) imaging for the tandem-and-ring applicator and X-ray computed tomography (CT) for the tandem-and-ovoids applicator (based on applicator compatibility)) to assess alignment. Four spherical fiducials were used to calculate target registration errors (TREs), providing a metric for validating registrations, where TREs were computed using root-mean-square distances to describe the alignment of manually identified corresponding fiducials. An analysis of variance (ANOVA) was used to identify statistically significant differences (p < 0.05) between the TREs for the three abdominal thicknesses for each applicator type. As an additional indicator of geometric accuracy, the bladder was segmented in the 3D US and corresponding MR/CT images, and volumetric differences and Dice similarity coefficients (DSCs) were calculated. RESULTS: For both applicator types, the combination of 3D TRUS with 3D TAUS images allowed image information obscured by the shadowing artifacts under single imaging perspectives to be recovered. For the tandem-and-ring applicator, the mean ± one standard deviation (SD) TREs from the images with increasing thicknesses were 1.37 ± 1.35 mm, 1.84 ± 1.22 mm, and 1.60 ± 1.00 mm. Similarly, for the tandem-and-ovoids applicator, the mean ± SD TREs from the images with increasing thicknesses were 1.37 ± 0.35 mm, 1.95 ± 0.90 mm, and 1.61 ± 0.76 mm. No statistically significant difference was detected in the TREs for the three thicknesses for either applicator type. The mean volume differences for the bladder segmentations were 3.14% and 2.33% and mean DSCs were 87.8% and 87.7% for the tandem-and-ring and tandem-and-ovoids applicators, respectively. CONCLUSIONS: In this proof-of-concept study, we demonstrated the feasibility of fusing 3D TRUS and 3D TAUS images based on the geometry of tandem-and-ring and tandem-and-ovoids applicators. This represents a step toward an accessible and low-cost 3D imaging method for gynecologic brachytherapy, with the potential to extend this approach to other intracavitary configurations and hybrid applicators.
Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.
How this classification was reachedexpand
Full frame machine prediction
Teacher imitationNot calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.
Distilled classifier scores by category (both heads)
| Category | Codex | Gemma |
|---|---|---|
| Metaresearch | 0.001 | 0.002 |
| Meta-epidemiology (narrow) | 0.001 | 0.001 |
| Meta-epidemiology (broad) | 0.000 | 0.001 |
| Bibliometrics | 0.001 | 0.000 |
| Science and technology studies | 0.000 | 0.000 |
| Scholarly communication | 0.001 | 0.001 |
| Open science | 0.001 | 0.001 |
| Research integrity | 0.001 | 0.000 |
| Insufficient payload (model declined to judge) | 0.001 | 0.000 |
Machine scores (provisional)
The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.
Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.
score_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from itClassification
machine, unvalidatedMachine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.
How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".