MétaCan
Menu
Back to cohort
Record W3197176510 · doi:10.1002/acm2.13385

Mathematical estimation of half‐value layer thicknesses

2021· article· en· W3197176510 on OpenAlexaboutno aff
Farid H. Omoumi, Xizeng Wu, Muhammad U. Ghani, Molly D. Wong, Yuhua Li, Hong Liu

Bibliographic record

VenueJournal of Applied Clinical Medical Physics · 2021
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsnot available
FundersNational Cancer Institute
KeywordsEstimationValue (mathematics)Layer (electronics)StatisticsMathematicsMaterials scienceEngineeringComposite material

Abstract

fetched live from OpenAlex

OBJECTIVE: The objective of this article is to introduce a simplified and swift method to satisfactorily estimate the half-value layers (HVL), quarter-value layer (QVL), and tenth-value layer (TVL) from the x-ray spectra emitted by any diagnostic radiology or kV radiotherapy x-ray tubes. METHODS: A CdTe x-ray and Gamma detector (X-123 CdTe, AmpTek Inc.) is used to measure the x-ray spectra at four different x-ray energies (low, mid, high energy x-rays) with different external filtering. The software "SpekCalc GUI" (Developed in McGill University, Montreal, Canada) is also used to obtain the simulated x-ray spectra. Both measured and simulated spectra are used to compute the HVL thicknesses of Aluminum by a mathematical method presented in this article. Next, the HVL thicknesses for corresponding tube potentials are also measured by calibrated ionization chamber and varying thicknesses of aluminum plates. Finally, the computed and measured HVL, QVL, and TVL thicknesses are compared to evaluate the efficacy of the presented method. RESULTS: The results show acceptable concordance between computed and measured quantities. The disagreement rates between measured HVL and the values derived mathematically from the x-ray spectra are 10 to 90 micrometers of Aluminum at tube potentials of 31 kV to 120 kV. As it is shown, a negligible discrepancy is observed between the analytical estimation and the experimental assessments. CONCLUSION: The HVL is an essential component in the evaluation of the quality of an x-ray beam. However, its measurement could occasionally be challenging, time-consuming, or uncertain due to some technical difficulties. Although the scope of this study is not to undermine the value of conventional and widely accepted practice to determine the HVL thickness, the introduced method provides the fast, more convenient, and comparably reliable technique to estimate the HVL, QVL, and TVL by employing the given x-ray spectrum.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.004
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Simulation or modeling · Consensus signal: none
GenreCandidate signal: Methods · Consensus signal: Methods
Teacher disagreement score0.003
Threshold uncertainty score0.009

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.004
Meta-epidemiology (narrow)0.0010.001
Meta-epidemiology (broad)0.0000.001
Bibliometrics0.0020.001
Science and technology studies0.0000.001
Scholarly communication0.0010.002
Open science0.0020.001
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0030.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.030
GPT teacher head0.385
Teacher spread0.355 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designSimulation or modeling
Domainnot available
GenreMethods

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations14
Published2021
Admission routes1
Has abstractyes

Explore more

Same venueJournal of Applied Clinical Medical PhysicsSame topicAdvanced Radiotherapy TechniquesFrench-language works237,207