Abstract

fetched live from OpenAlex

BACKGROUND: Respiratory rate (RR) is arguably the most important vital sign to detect clinical deterioration. Change in RR can also, for example, be associated with the onset of different diseases, opioid overdoses, intense workouts, or mood. However, unlike for most other vital parameters, an easy and accurate measuring method is lacking. OBJECTIVE: This study aims to validate the radar-based sleep monitor, Somnofy, for measuring RRs and investigate whether events affecting RR can be detected from personalized baselines calculated from nightly averages. METHODS: First, RRs from Somnofy for 37 healthy adults during full nights of sleep were extensively validated against respiratory inductance plethysmography. Then, the night-to-night consistency of a proposed filtered average RR was analyzed for 6 healthy participants in a pilot study in which they used Somnofy at home for 3 months. RESULTS: Somnofy measured RR 84% of the time, with mean absolute error of 0.18 (SD 0.05) respirations per minute, and Bland-Altman 95% limits of agreement adjusted for repeated measurements ranged from -0.99 to 0.85. The accuracy and coverage were substantially higher in deep and light sleep than in rapid eye movement sleep and wake. The results were independent of age, sex, and BMI, but dependent on supine sleeping position for some radar orientations. For nightly filtered averages, the 95% limits of agreement ranged from -0.07 to -0.04 respirations per minute. In the longitudinal part of the study, the nightly average was consistent from night to night, and all substantial deviations coincided with self-reported illnesses. CONCLUSIONS: RRs from Somnofy were more accurate than those from any other alternative method suitable for longitudinal measurements. Moreover, the nightly averages were consistent from night to night. Thus, several factors affecting RR should be detectable as anomalies from personalized baselines, enabling a range of applications. More studies are necessary to investigate its potential in children and older adults or in a clinical setting.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.003
metaresearch head score (Gemma)0.005
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.003
Threshold uncertainty score0.018

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0030.005
Meta-epidemiology (narrow)0.0010.000
Meta-epidemiology (broad)0.0000.001
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0010.000
Research integrity0.0010.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.039
GPT teacher head0.285
Teacher spread0.246 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations18
Published2022
Admission routes1
Has abstractyes

Explore more

Same venueJMIR Biomedical EngineeringSame topicNon-Invasive Vital Sign MonitoringFrench-language works237,207