MétaCan
Menu
Back to cohort
Record W4296129422 · doi:10.1016/j.jscai.2022.100449

First-in-Human Experience With the ModulHeart Device for Mechanical Circulatory Support and Renal Perfusion

2022· article· en· W4296129422 on OpenAlexaff
Gabriel Georges, François Trudeau, Jade Doucet-Martineau, Maxime Rochon, Jeannot Potvin, Adrián Ebner, Philippe Généreux

Bibliographic record

VenueJournal of the Society for Cardiovascular Angiography & Interventions · 2022
Typearticle
Languageen
FieldEngineering
TopicMechanical Circulatory Support Devices
Canadian institutionsCentre Hospitalier de l’Université de MontréalWalsh Medical Devices (Canada)
Fundersnot available
KeywordsCirculatory systemPerfusionMedicineCardiologyInternal medicine

Abstract

fetched live from OpenAlex

Background: ModulHeart (Puzzle Medical Devices Inc) is a modular device providing hemodynamic support through 3 endovascular pumps inserted in series and assembled in parallel into a self-expandable anchor implanted in the descending aorta. The current study evaluates the feasibility and safety of cardiorenal support with ModulHeart among patients undergoing high-risk percutaneous coronary intervention (PCI). Methods: This study was a prospective, single-center, first-in-human study. The primary end point was procedural success, defined as successful delivery, function, and removal of the ModulHeart device. Secondary end points included pump hemodynamics, cardiac hemodynamics, and urine output. Results: On June 28 and 29, 2022, 4 patients were enrolled and underwent high-risk PCI with ModulHeart implanted via transfemoral approach. All 4 patients achieved procedural success. The mean delivery time was 8 minutes, the mean support time was 49 minutes, and the mean pump removal time was 7 minutes. The mean pressure gradient across the pump was 5 ± 2 mm Hg. Under ModulHeart support, cardiac index increased by 25%, central venous pressure decreased by 37%, and left ventricular end-diastolic pressure decreased by 78%. Urine output increased by ∼9-fold after 15 minutes of support. No device malfunction or procedural or device-related adverse events occurred. There was no evidence of pump thrombosis. All 4 patients were alive at 30 days. Conclusions: This first-in-human study demonstrated the feasibility and safety of cardiorenal support with ModulHeart among patients undergoing high-risk PCI. ModulHeart demonstrated significant improvement in cardiac output, left ventricular end-diastolic pressure, and urine output. Future studies are planned to assess outcomes associated with ModulHeart support in patients with heart failure.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.002
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesMeta-epidemiology (broad)
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Not applicable · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.705
Threshold uncertainty score0.956

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0020.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.054
Bibliometrics0.0000.000
Science and technology studies0.0010.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.020
GPT teacher head0.246
Teacher spread0.226 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

Study designNot applicable
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations13
Published2022
Admission routes1
Has abstractyes

Explore more

Same venueJournal of the Society for Cardiovascular Angiography & InterventionsSame topicMechanical Circulatory Support DevicesFrench-language works237,207