MétaCan
Menu
Back to cohort
Record W4387297168 · doi:10.1111/aor.14655

Organ blood flow assessment with the <scp>ModulHeart</scp> cardiorenal support device

2023· article· en· W4387297168 on OpenAlexaff
Gabriel Georges, François Trudeau, Jeannot Potvin, Daniel Burkhoff, Philippe Généreux

Bibliographic record

VenueArtificial Organs · 2023
Typearticle
Languageen
FieldEngineering
TopicMechanical Circulatory Support Devices
Canadian institutionsCentre Hospitalier de l’Université de MontréalWalsh Medical Devices (Canada)
Fundersnot available
KeywordsMedicineCardiologyPerfusionInternal medicineHemodynamicsBlood flowBlood pressureRenal circulationVentricleDescending aortaAortaCardiac outputAortic pressureCerebral perfusion pressureRenal blood flowCerebral blood flowPulse pressureCoronary perfusion pressureAnesthesiaResuscitation

Abstract

fetched live from OpenAlex

BACKGROUND: ModulHeart (Puzzle Medical Devices Inc) is a novel percutaneous flow entrainment pump anchored in the descending aorta. The current study evaluates the hemodynamic effect of ModulHeart support and its impact on cerebral, myocardial, and renal blood flow. METHODS: ModulHeart was implanted in the descending aorta of four healthy calves. A ramp protocol (2000 RPM increments) was performed with the pump operating at five different speeds from 14 000 to 22 000 RPM. For each speed, pressures proximal and distal to the pump, and right heart catheterization measurements were recorded. Stable-isotope labeled microspheres were injected in the left ventricle to evaluate organ perfusion. RESULTS: Thermodilution cardiac output increased by 23% at 22 000 RPM. Greater pump speeds resulted in greater pump gradients, up to 10 mm Hg in mean arterial pressure at 22 000 RPM, without significant reduction of proximal perfusion pressures. Arterial pulse pressure remained stable at all speeds. ModulHeart was not associated with a reduction in cerebral or myocardial blood flow at any speed. Renal cortical and medullary blood flow increased by up to 50% and 40%, respectively. CONCLUSION: The ModulHeart device implanted in the descending aorta of healthy calves resulted in significant arterial pressure gradients and preserved pulse pressure. Greater pump speeds translated into greater increases in renal blood flow, with no decrease in cerebral or myocardial perfusion.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesInsufficient payload (model declined to judge)
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.145
Threshold uncertainty score0.999

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0010.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.001
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.002

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.017
GPT teacher head0.231
Teacher spread0.214 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations2
Published2023
Admission routes1
Has abstractyes

Explore more

Same venueArtificial OrgansSame topicMechanical Circulatory Support DevicesFrench-language works237,207