A-349 Evaluation of <i>i-STAT</i>® Point of Care Blood Gas Cartridges and Competitor Blood Gas Devices Against Reference Standard for <i>P</i>CO2 and <i>P</i>O2

2024· article· en· W4403048869 on OpenAlexaff
Craig W. Beattie, Linda Thibodeau

Bibliographic record

VenueClinical Chemistry · 2024
Typearticle
Languageen
FieldMedicine
TopicHemodynamic Monitoring and Therapy
Canadian institutionsAbbott (Canada)
Fundersnot available
KeywordsCartridgePoint of carePoint-of-care testingWhole bloodChromatographyMedicineChemistryImmunologyMaterials sciencePathology

Abstract

fetched live from OpenAlex

Abstract Background The i-STAT System provides laboratory quality results in minutes at the patient’s bedside. Accurate and rapid test results are critical for clinical decision making in the presence of blood gas disorders, including oxygenation and acid-base status, where the partial pressure oxygen (PO2) and partial pressure carbon dioxide (PCO2) are needed. The purpose of this study was to compare the analytical performance of the PO2 and PCO2 tests in the i-STAT G3+ and i-STAT CG8+ cartridges to the theoretical PO2 and PCO2 in prepared reference standards. Two other blood gas instruments, a benchtop device and a laboratory device were also compared to the reference PO2 and PCO2. Methods Venous whole blood samples were contrived using saturation tonometry with U.S. NIST (National Institute of Science and Technology) traceable gas tanks to prepare the reference standards, which were value assigned to theoretical PO2 or PCO2 levels based on the molar composition of the gas mixture used. Eleven levels spanning the reportable range of each PO2 (5 mmHg - 800 mmHg) and PCO2 (5 mmHg - 130 mmHg) were prepared and tested in duplicate on the i-STAT G3+ and i-STAT CG8+ cartridges, benchtop device, and the laboratory device. Passing-Bablok linear regression analysis was performed to evaluate the slope and correlation coefficient, comparing each blood gas device against the reference PO2 and PCO2 values. Study designs followed CLSI (Clinical and Laboratory Standards Institute) EP09C-ED3:2018, Measurement Procedure Comparison and Bias Estimation using Patient Samples, 3rd Edition. Passing-Bablok regression analysis was also performed for the i-STAT cartridges and benchtop device against the laboratory device. Results The regression analysis was performed against the reference standards. For PO2, slopes for the i-STAT cartridges ranged from 0.93 - 0.97, the slope for the benchtop device was 0.99, and the slope for the laboratory device was 0.97. Correlation coefficients for all devices were 1.00. For PCO2, slopes for the i-STAT cartridges ranged from 0.98 - 1.02, the slope for the benchtop device was 0.85, and the slope for the laboratory device was 1.01. Correlation coefficients were 1.00 for the i-STAT cartridges and laboratory device, and 0.99 for the benchtop device. The regression analysis was also performed against the laboratory device. For PO2, slopes for the i-STAT cartridges ranged from 0.96 - 1.00, and the slope for the benchtop device was 1.02. Correlation coefficients for all devices were 1.00. For PCO2, slopes for the i-STAT cartridges ranged from 0.97 - 1.01, and the slope for the benchtop device was 0.85. Correlation coefficients were 1.00 for the i-STAT cartridges and 0.99 for the benchtop device. Conclusions The study demonstrated that the i-STAT G3+ and i-STAT CG8+ cartridges used with the i-STAT System were shown to provide laboratory quality results within 2 minutes, showing good agreement to both reference standards and laboratory quality devices. For PCO2, the i-STAT System was shown to have better agreement to both theoretical reference standards and the laboratory device than the benchtop device. These studies were funded by Abbott Laboratories.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.001
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.461
Threshold uncertainty score0.671

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0010.001
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.045
GPT teacher head0.381
Teacher spread0.336 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations2
Published2024
Admission routes1
Has abstractyes

Explore more

Same venueClinical ChemistrySame topicHemodynamic Monitoring and TherapyFrench-language works237,207