Correcting scan‐to‐scan response variability for a radiochromic film‐based reference dosimetry system

2015· article· en· W1559985845 on OpenAlexafffund
David F. Lewis, Slobodan Dević

Bibliographic record

VenueMedical Physics · 2015
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsMcGill UniversityJewish General Hospital
FundersNatural Sciences and Engineering Research Council of Canada
KeywordsScannerDosimetryMaterials scienceOpticsNuclear medicineStandard deviationBiomedical engineeringPhysicsMedicineMathematics

Abstract

fetched live from OpenAlex

PURPOSE: In radiochromic film dosimetry systems, measurements are usually obtained from film images acquired on a CCD-based flatbed scanner. The authors investigated factors affecting scan-to-scan response variability leading to increased dose measurement uncertainty. METHODS: The authors used flatbed document scanners to repetitively scan EBT3 radiochromic films exposed to doses 0-1000 cGy, together with three neutral density filters and three blue optical filters. Scanning was performed under two conditions: scanner lid closed and scanner lid opened/closed between scans. The authors also placed a scanner in a cold room at 9 °C and later in a room at 22 °C and scanned EBT3 films to explore temperature effects. Finally, the authors investigated the effect of altering the distance between the film and the scanner's light source. RESULTS: Using a measurement protocol to isolate the contribution of the CCD and electronic circuitry of the scanners, the authors found that the standard deviation of response measurements for the EBT3 film model was about 0.17% for one scanner and 0.09% for the second. When the lid of the first scanner was opened and closed between scans, the average scan-to-scan difference of responses increased from 0.12% to 0.27%. Increasing the sample temperature during scanning changed the RGB response values by about -0.17, -0.14, and -0.05%/°C, respectively. Reducing the film-to-light source distance increased the RBG response values about 1.1, 1.3, and 1.4%/mm, respectively. The authors observed that films and film samples were often not flat with some areas up to 8 mm away from the scanner's glass window. CONCLUSIONS: In the absence of measures to deal with the response irregularities, each factor the authors investigated could lead to dose uncertainty >2%. Those factors related to the film-to-light source distance could be particularly impactful since the authors observed many instances where the curl of film samples had the potential to cause dose uncertainty in excess of 5%. Two expedients will eliminate the uncertainties: a transparent sheet (preferably glass) placed over the scanned film keeps the film-to-light source distance constant, and an EBT3 reference film included in all scans provides correction factors for measured response values.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.006
metaresearch head score (Gemma)0.015
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: none
Teacher disagreement score0.006
Threshold uncertainty score0.032

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0060.015
Meta-epidemiology (narrow)0.0010.001
Meta-epidemiology (broad)0.0010.001
Bibliometrics0.0010.001
Science and technology studies0.0010.001
Scholarly communication0.0010.001
Open science0.0020.001
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0010.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.025
GPT teacher head0.319
Teacher spread0.294 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations58
Published2015
Admission routes2
Has abstractyes

Explore more

Same venueMedical PhysicsSame topicAdvanced Radiotherapy TechniquesFrench-language works237,207