Simulation study on the heating of the surrounding anatomy during transurethral ultrasound prostate therapy: A 3D theoretical analysis of patient safety

2010· article· en· W2037852391 on OpenAlexafffund
Mathieu Burtnyk, Rajiv Chopra, Michael Bronskill

Bibliographic record

VenueMedical Physics · 2010
Typearticle
Languageen
FieldEngineering
TopicUltrasound and Hyperthermia Applications
Canadian institutionsHealth Sciences CentreUniversity of TorontoSunnybrook Health Science Centre
FundersNatural Sciences and Engineering Research Council of CanadaCanadian Institutes of Health ResearchTerry Fox FoundationOntario Research Foundation
KeywordsNeurovascular bundleUltrasoundProstateRectumMedicineTransducerProstate cancerBiomedical engineeringRadiologyMaterials scienceAnatomySurgeryCancerAcoustics

Abstract

fetched live from OpenAlex

PURPOSE: MRI-guided transurethral ultrasound therapy can generate highly accurate volumes of thermal coagulation conforming to 3D human prostate geometries. This work simulated, quantified, and evaluated the thermal impact of these treatments on the rectum, pelvic bone, neurovascular bundles (NVBs), and urinary sphincters because damage to these structures can lead to complications. METHODS: Twenty 3D anatomical models of prostate cancer patients were used with detailed bioacoustic simulations incorporating an active feedback algorithm which controlled a rotating, planar ultrasound transducer (17, 4 x 3 mm2 elements, 10 W(acoustic)/cm2). Heating of the adjacent surrounding anatomy was evaluated at 4.7, 9.7, and 14.2 MHz using thermal tolerances reported in literature. RESULTS: Heating of the rectum posed the most important safety concern, influenced largely by the water temperature of an endorectal cooling device (ECD); depending on anatomy, temperatures of 7-37 degrees C were required to limit potential damage to less than 10 mm3 on the outer 1 mm layer of the rectal wall. Heating of the pelvic bone could be important at 4.7 MHz. A smaller sized ECD or a higher ultrasound frequency in sectors where the bone was less than 10 mm from the prostate reduced heating in all cases below the threshold for irreversible damage. Heating of the NVB was significant in 75% of the patient models in the absence of treatment planning; this proportion was reduced to 5% by increasing treatment margins up to 4 mm. To avoid damaging the urinary sphincters, the transducer should be positioned at least 2-4 mm from the sphincters, depending on the transurethral cooling temperature. CONCLUSIONS: Simulations show that MRI-guided transurethral therapy can treat the prostate accurately, but in the absence of treatment planning, some thermal impact can be predicted on the surrounding anatomy. Treatment planning strategies have been developed, which reduce thermal injury to the surrounding anatomy.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Simulation or modeling · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.288
Threshold uncertainty score0.292

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.001
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0000.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.007
GPT teacher head0.247
Teacher spread0.240 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designSimulation or modeling
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations32
Published2010
Admission routes2
Has abstractyes

Explore more

Same venueMedical PhysicsSame topicUltrasound and Hyperthermia ApplicationsFrench-language works237,207