TU‐E‐108‐04: Preliminary Measurements Towards Radiation Therapy Involving Dynamic Couch Trajectories

2013· article· en· W2047657667 on OpenAlexaff
Joel Mullins, Jan Seuntjens, F DeBlois, Alasdair Syme

Bibliographic record

VenueMedical Physics · 2013
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsMcGill University
Fundersnot available
KeywordsIonization chamberTruebeamEclipseNuclear medicinePhysicsOpticsIonMaterials scienceLinear particle acceleratorBeam (structure)Medicine

Abstract

fetched live from OpenAlex

Purpose: To assess the geometrical accuracy of treatment deliveries involving dynamic couch trajectories and to evaluate the accuracy of DMLC dose calculations in Eclipse at nonstandard SSDs. Methods: Eleven DMLC IMRT plans were recalculated as verification plans in Eclipse (AAA 10.0.28) using the collapsed gantry method, at SSDs between 85‐135 cm. Ion chamber and film measurements were performed to compare planned and delivered dose. DLG and MLC transmission were also measured as a function of SSD. Simple circular couch trajectories that simulated isocentric treatment at an SAD of 95 cm were evaluated using film and ion chamber measurements. Trajectories were implemented using Varian TrueBeam Developers Mode. Results: The relative agreement between Eclipse‐calculated dose and ion chamber measurements decreased systematically by an average of 1.5±0.3 % between 85 and 135 cm SSD. Film measurements were consistent with the ion chamber results. Following corrections for the geometric projection of the DLG, no dependence on SSD was observed, and the dependence of MLC transmission on SSD was negligible. These results are both consistent with Eclipse calculations. The standard error of ion chamber current measurements was 0.19 % for the dynamic couch trajectory at 95 cm SAD, compared to 0.17 % for static couch position. From the film measurements, the FWHM of the field delivered with dynamic couch motion was 94.6 % of the FWHM with static couch position, matching the predicted geometrical minification. Conclusion: Experiments were performed to investigate aspects of radiation therapy involving dynamic couch trajectories. The relative agreement between calculated and measured dose was found to systematically decrease with SSD, while DLG and MLC transmission were independent of SSD. Isocenter positional accuracy with dynamic couch motion was demonstrated through the consistency of ion chamber current measurements and observation of the expected decrease in field size at a shortened SAD.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.003
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.005
Threshold uncertainty score0.009

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.003
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0030.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.019
GPT teacher head0.288
Teacher spread0.269 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2013
Admission routes1
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced Radiotherapy Techniques→French-language works237,207→