Sci—Wed PM: Delivery—07: Dosimetric Characterization of a New Pixel Ionization Chamber Array

2009· article· en· W2069908654 on OpenAlexaff
Sankar Venkataraman, KE Malkoske, Martin Bach Jensen, G Asuni, KD Nakonechny, Anita Berndt, BMC McCurdy

Bibliographic record

VenueMedical Physics · 2009
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsCancerCare Manitoba
Fundersnot available
KeywordsIonization chamberDosimetryMonitor unitOpticsDose profileLinear particle acceleratorNuclear medicineMaterials scienceDetectorCompassLinearityReproducibilityBeam (structure)IonizationPhysicsIonChemistryMedicine

Abstract

fetched live from OpenAlex

A new ionization chamber array developed by IBA Dosimetry (Germany) is evaluated at our centre. The dosimetric characteristics of this device such as dose and dose rate dependence, stability, energy dependence, profiles and error detection ability were studied. COMPASS is an array with 1600 parallel plate ion chambers. Each chamber is 3.8 mm in diameter and 2mm in height spaced 6.5 mm apart. The array is mounted onto the wedge slot of the Varian Clinac 2100iX (Varian Medical Systems, Palo Alto, CA) linear accelerator. Linearity of the detector with dose and response to dose rate were done using Clinac 2100iX 6 MV beam. The detector showed excellent linearity response with dose and a variation of <0.2% with dose rates. Short term and long term reproducibility results showed a standard deviation of 0.05%. Energy dependence of COMPASS was evaluated by response of the device to beam energies ranging from to 18 MV X‐rays. , determined by COMPASS increases with energy as opposed to a decreased response with energy of a thimble ion chamber. Profiles measured with COMPASS were similar to Markus (plane parallel) ion chamber for field sizes . The chamber array showed increased dose response off‐axis for larger fields. The penumbra measured by COMPASS was 3.14± 0.03 mm vs. 1.25± 0.02 mm for film measurements for a range of field sizes. A leaf gap error of 1 mm was detected by the device using an IMRT test field. Research partially sponsored by IBA Dosimetry.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.001
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.007
Threshold uncertainty score0.023

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.001
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0010.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0070.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.010
GPT teacher head0.263
Teacher spread0.253 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2009
Admission routes1
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced Radiotherapy Techniques→French-language works237,207→