MétaCan
Menu
Back to cohort

Artificial Human Corneas

2002· review· en· W2074696012 on OpenAlexaff
May Griffith, M. Obaidul Hakim, Shigeto Shimmura, Mitchell A. Watsky, Fengfu Li, David Carlsson, Charles J. Doillon, Masatsugu Nakamura, Erik J. Suuronen, Naoshi Shinozaki, Katsuhiko Nakata, Heather Sheardown

Bibliographic record

VenueCornea · 2002
Typereview
Languageen
FieldMedicine
TopicCorneal Surgery and Treatments
Canadian institutionsUniversity of Ottawa
Fundersnot available
KeywordsCorneaBiocompatibilityBiomedical engineeringTransplantationCorneal transplantationMaterials scienceTissue engineeringStem cellScaffoldSurgeryOphthalmologyMedicineBiologyCell biology

Abstract

fetched live from OpenAlex

PURPOSE: To review the development of artificial corneas (prostheses and tissue equivalents) for transplantation, and to provide recent updates on our tissue-engineered replacement corneas. METHODS: Modified natural polymers and synthetic polymers were screened for their potential to replace damaged portions of the human cornea or the entire corneal thickness. These polymers, combined with cells derived from each of the three main corneal layers or stem cells, were used to develop artificial corneas. Functional testing was performed in vitro. Trials of biocompatibility and immune and inflammatory reactions were performed by implanting the most promising polymers into rabbit corneas. RESULTS: Collagen-based biopolymers, combined with synthetic crosslinkers or copolymers, formed effective scaffolds for developing prototype artificial corneas that could be used as tissue replacements in the future. We have previously developed an artificial cornea that mimicked key morphologic and functional properties of the human cornea. The addition of synthetic polymers increased its toughness as it retained transparency and low light scattering, making the matrix scaffold more suitable for transplantation. These new composites were implanted into rabbits without causing any acute inflammation or immune response. We have also fabricated full-thickness composites that can be fully sutured. However, the long-term effects of these artificial corneas need to be evaluated. CONCLUSIONS: Novel tissue-engineered corneas that comprise composites of natural and synthetic biopolymers together with corneal cell lines or stem cells will, in the future, replace portions of the cornea that are damaged. Our results provide a basis for the development of both implantable temporary and permanent corneal replacements.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.001
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Not applicable · Consensus signal: none
GenreCandidate signal: Review · Consensus signal: Review
Teacher disagreement score0.009
Threshold uncertainty score0.031

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.001
Meta-epidemiology (narrow)0.0010.000
Meta-epidemiology (broad)0.0000.001
Bibliometrics0.0010.001
Science and technology studies0.0000.001
Scholarly communication0.0010.001
Open science0.0010.001
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0090.005

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.253
GPT teacher head0.403
Teacher spread0.150 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designNot applicable
Domainnot available
GenreReview

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations115
Published2002
Admission routes1
Has abstractyes

Explore more

Same venueCorneaSame topicCorneal Surgery and TreatmentsFrench-language works237,207