MétaCan
Menu
← Back to cohort
Record W2090499975 · doi:10.1118/1.3613015

MO‐F‐214‐02: A New, High Resolution 2D Dose Measurement Device Based on a Few Scintillating Fibers and Tomographic Reconstruction

2011· article· en· W2090499975 on OpenAlexaff
M Goulet, L Gingras, Louis Archambault, Luc Beaulieu

Bibliographic record

VenueMedical Physics · 2011
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsUniversité LavalHôtel-Dieu de Québec
Fundersnot available
KeywordsImaging phantomOpticsDosimeterPixelImage resolutionMaterials scienceIterative reconstructionPhysicsField of viewProjection (relational algebra)Reconstruction algorithmNuclear medicineRadiationComputer scienceMedicineArtificial intelligenceAlgorithm

Abstract

fetched live from OpenAlex

Purpose: To develop and validate a novel type of 2D dosimeter based on the tomographic reconstruction of the dose projections obtained using long scintillating fibers. Methods: 50 parallel scintillating fibers (diameter, 1mm; length, 6 to 20cm) were aligned in a 30cm diameter cylindrical masonite phantom with a 90cm source‐to‐surface distance and a 100cm source‐to‐fibers distance. The fibers were disposed so that the effective detection area of the scintillating fibers was a 20cm diameter disk. Both ends of each scintillating fiber were coupled to clear optical fibers to enable light collection by a single CCD camera using an f/2, 50mm focal length lens. 7 IMRT segments and 2 square fields were acquired using 18 projections over a 170 degrees rotation of the device. Dose reconstructions were conducted using a total‐variation minimization reconstruction algorithm. 8 monitoring units were programmed for each projection and the reconstructed dose grid pixel resolution was set to 1×1mm,2,. Results: Using a non‐optimized algorithm on a 2GHz CPU, each reconstruction was performed in less than 6 minutes. 3%/3mm gamma tests conducted between the reconstructed IMRT dose distributions and the dose calculated with the treatment planning system Pinnacle,3, were on average successful for 99.6% of the dose pixels for the region over 10% of the maximum dose. For both square fields and the whole summed IMRT field, 100% of the dose pixels were successful to the gamma test. Conclusions: Using tomographic reconstruction on the projections acquired with rotating scintillating fibers, one is able to perform 2D dosimetry of simple and IMRT fields with great accuracy and resolution using only a limited number of scintillating fibers. The underlying concept of tomographic dosimetry and the small number of fibers needed to reconstruct a given 2D dose distribution offer a world of new dosimetric possibility, both applicable to 2D and 3D dosimetry.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.001
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: none
Teacher disagreement score0.002
Threshold uncertainty score0.008

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.001
Meta-epidemiology (narrow)0.0010.001
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0010.001
Science and technology studies0.0000.000
Scholarly communication0.0010.001
Open science0.0010.001
Research integrity0.0010.000
Insufficient payload (model declined to judge)0.0020.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.029
GPT teacher head0.256
Teacher spread0.227 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations3
Published2011
Admission routes1
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced Radiotherapy Techniques→French-language works237,207→