MétaCan
Menu
Back to cohort
Record W2091474364 · doi:10.1016/j.jbi.2014.04.009

The use of fault reporting of medical equipment to identify latent design flaws

2014· article· en· W2091474364 on OpenAlexaff
Christopher Flewwelling, Anthony Easty, Kim J. Vicente, Joseph A Cafazzo

Bibliographic record

VenueJournal of Biomedical Informatics · 2014
Typearticle
Languageen
FieldHealth Professions
TopicQuality and Safety in Healthcare
Canadian institutionsUniversity Health NetworkUniversity of Toronto
Fundersnot available
KeywordsUsabilityHeuristic evaluationUsability labCognitive walkthroughUsability inspectionUsability engineeringComputer scienceSystem usability scalePluralistic walkthroughEngineeringHuman–computer interaction

Abstract

fetched live from OpenAlex

BACKGROUND AND PURPOSE: Poor device design that fails to adequately account for user needs, cognition, and behavior is often responsible for use errors resulting in adverse events. This poor device design is also often latent, and could be responsible for "No Fault Found" (NFF) reporting, in which medical devices sent for repair by clinical users are found to be operating as intended. Unresolved NFF reports may contribute to incident under reporting, clinical user frustration, and biomedical engineering technologist inefficacy. This study uses human factors engineering methods to investigate the relationship between NFF reporting frequency and device usability. MATERIAL AND METHODS: An analysis of medical equipment maintenance data was conducted to identify devices with a high NFF reporting frequency. Subsequently, semi-structured interviews and heuristic evaluations were performed in order to identify potential usability issues. Finally, usability testing was conducted in order to validate that latent usability related design faults result in a higher frequency of NFF reporting. RESULTS: The analysis of medical equipment maintenance data identified six devices with a high NFF reporting frequency. Semi-structured interviews, heuristic evaluations and usability testing revealed that usability issues caused a significant portion of the NFF reports. Other factors suspected to contribute to increased NFF reporting include accessory issues, intermittent faults and environmental issues. Usability testing conducted on three of the devices revealed 23 latent usability related design faults. CONCLUSIONS: These findings demonstrate that latent usability related design faults manifest themselves as an increase in NFF reporting and that devices containing usability related design faults can be identified through an analysis of medical equipment maintenance data.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.023
metaresearch head score (Gemma)0.155
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: Observational
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.023
Threshold uncertainty score0.120

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0230.155
Meta-epidemiology (narrow)0.0010.000
Meta-epidemiology (broad)0.0010.001
Bibliometrics0.0120.004
Science and technology studies0.0010.001
Scholarly communication0.0020.003
Open science0.0010.002
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0010.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.397
GPT teacher head0.534
Teacher spread0.138 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations29
Published2014
Admission routes1
Has abstractyes

Explore more

Same venueJournal of Biomedical InformaticsSame topicQuality and Safety in HealthcareFrench-language works237,207