Record W2093758774 · doi:10.1118/1.3669487

Dose homogeneity specification for reference dosimetry of nonstandard fields

2011· article· en· W2093758774 on OpenAlexafffund
Eunah Chung, E Soisson, Jan Seuntjens

Bibliographic record

VenueMedical Physics · 2011
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsMcGill University Health CentreMcGill UniversityMontreal General Hospital
FundersNatural Sciences and Engineering Research Council of CanadaInternational Atomic Energy Agency
KeywordsDosimetryIonization chamberImaging phantomPhysicsPercentage depth dose curveNuclear medicineAbsorbed doseDose profileIonizationWater equivalentOpticsMaterials scienceMedicineIon

Abstract

fetched live from OpenAlex

PURPOSE: To investigate the sensitivity of the plan-class specific correction factor to dose distributions in composite nonstandard field dosimetry. METHODS: A cylindrical water-filled PMMA phantom was constructed at the center of which reference absorbed dose could be measured. Ten different TomoTherapy(®)-based IMRT fields were created on the CT images of the phantom. The dose distribution for each IMRT field was estimated at the position of a radiation detector or ionization chamber. The dose in each IMRT field normalized to that in a reference 10 × 10 cm(2) field was measured using a PTW micro liquid ion chamber. Based on the new dosimetry formalism, a plan-class specific correction factor k(Q(pcsr),Q) (f(pcsr),f(ref)) for each field was measured for two Farmer-type chambers, Exradin A12 and NE2571, as well as for a smaller Exradin A1SL chamber. The dependence of the measured correction factor on parameters characterizing dose distribution was analyzed. RESULTS: Uncertainty on the plan-class specific correction factor measurement was in the range of 0.3%-0.5% and 0.3%-0.8% for the Farmer-type chambers and the Exradin A1SL, respectively. When the heterogeneity of the central region of the target volume was less than 5%, the correction factor did not differ from unity by more than 0.7% for the three air-filled ionization chambers. For more heterogeneous dose deliveries, the correction factor differed from unity by up to 2.4% for the Farmer-type chambers. For the Exradin A1SL, the correction factor was closer to unity due to the reduced effect of dose gradients, while it was highly variable in different IMRT fields because of a more significant impact of positioning uncertainties on the response of this chamber. CONCLUSIONS: The authors have shown that a plan-class specific correction factor can be specified as a function of plan evaluation parameters especially for Farmer-type chambers. This work provides a recipe based on quantifying dose distribution to accurately select air-filled ionization chamber correction factors for nonstandard fields.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame distilled prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.

metaresearch head score (Codex)0.000
metaresearch head score (Gemma)0.000
Version: codex-gemma-dda1882f352aValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: none
Teacher disagreement score0.961
Threshold uncertainty score0.641

Codex and Gemma teacher scores by category

CategoryCodexGemma
Metaresearch0.0000.000
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0010.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.043
GPT teacher head0.314
Teacher spread0.271 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one teacher head, not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations21
Published2011
Admission routes2
Has abstractyes

Explore more

Same venueMedical PhysicsSame topicAdvanced Radiotherapy TechniquesFrench-language works237,207