A Monte Carlo approach to validation of FFF VMAT treatment plans for the TrueBeam linac
Bibliographic record
Abstract
PURPOSE: To commission and benchmark a vendor-supplied (Varian Medical Systems) Monte Carlo phase-space data for the 6 MV flattening filter free (FFF) energy mode on a TrueBeam linear accelerator for the purpose of quality assurance of clinical volumetric modulated arc therapy (VMAT) treatment plans. A method for rendering the phase-space data compatible with BEAMnrc/DOSXYZnrc simulation software package is presented. METHODS: Monte Carlo (MC) simulations were performed to benchmark the TrueBeam 6 MV FFF phase space data that have been released by the Varian MC Research team. The simulations to benchmark the phase space data were done in three steps. First, the original phase space which was created on a cylindrical surface was converted into a format that was compatible with BEAMnrc. Second, BEAMnrc was used to create field size specific phase spaces located underneath the jaws. Third, doses were calculated with DOSXYZnrc in a water phantom for fields ranging from 1 × 1 to 40 × 40 cm(2). Calculated percent depth doses (PDD), transverse profiles, and output factors were compared with measurements for all the fields simulated. After completing the benchmarking study, three stereotactic body radiotherapy (SBRT) VMAT plans created with the Eclipse treatment planning system (TPS) were calculated with Monte Carlo. Ion chamber and film measurements were also performed on these plans. 3D gamma analysis was used to compare Monte Carlo calculation with TPS calculations and with film measurement. RESULTS: For the benchmarking study, MC calculated and measured values agreed within 1% and 1.5% for PDDs and in-field transverse profiles, respectively, for field sizes >1 × 1 cm(2). Agreements in the 80%-20% penumbra widths were better than 2 mm for all the fields that were compared. With the exception of the 1 × 1 cm(2) field, the agreement between measured and calculated output factors was within 1%. It is of note that excellent agreement in output factors for all field sizes including highly asymmetric fields was achieved without accounting for backscatter into the beam monitor chamber. For the SBRT VMAT plans, the agreement between Monte Carlo and ion chamber point dose measurements was within 1%. Excellent agreement between Monte Carlo, treatment planning system and Gafchromic film dose distribution was observed with over 99% of the points in the high dose volume passing the 3%, 3 mm gamma test. CONCLUSIONS: The authors have presented a method for making the Varian IAEA compliant 6 MV FFF phase space file of the TrueBeam linac compatible with BEAMnrc/DOSXYZnrc. After benchmarking the modified phase space against measurement, they have demonstrated its potential for use in MC based quality assurance of complex delivery techniques.
Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.
How this classification was reachedexpand
Full frame distilled prediction
Teacher imitationNot calibrated prevalence, not ground truth. Human validation pending. Learned from the 10,348 direct Codex labels and 10,348 direct Gemma labels. Candidate is the union of thresholded teacher heads; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels or direct frontier model labels.
Codex and Gemma teacher scores by category
| Category | Codex | Gemma |
|---|---|---|
| Metaresearch | 0.000 | 0.000 |
| Meta-epidemiology (narrow) | 0.000 | 0.000 |
| Meta-epidemiology (broad) | 0.000 | 0.000 |
| Bibliometrics | 0.000 | 0.000 |
| Science and technology studies | 0.000 | 0.000 |
| Scholarly communication | 0.000 | 0.000 |
| Open science | 0.000 | 0.000 |
| Research integrity | 0.000 | 0.000 |
| Insufficient payload (model declined to judge) | 0.000 | 0.000 |
Machine scores (provisional)
The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.
Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.
score_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from itClassification
machine, unvalidatedMachine predicted; a candidate call from one teacher head, not a consensus.
How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".