Validation of a new generation POCT glucose device with emphasis on aspects important for glycemic control in the hospital care

2012· article· en· W2126864963 on OpenAlexfundno aff
Snježana Kos, Arie van Meerkerk, Joke van der Linden, Theo Stiphout, Remi W. Wulkan

Bibliographic record

VenueClinical Chemistry and Laboratory Medicine (CCLM) · 2012
Typearticle
Languageen
FieldMedicine
TopicHyperglycemia and glycemic control in critically ill and hospitalized patients
Canadian institutionsnot available
FundersUniversity of Calgary
KeywordsGlycemicHematocritMedicinePoint-of-care testingCuvetteBiomedical engineeringInternal medicineInsulinPathology

Abstract

fetched live from OpenAlex

BACKGROUND: Point-of-care (POC) glucose devices are widely used for insulin-dosage decision-making although such an application is not always permitted. In this study, we have evaluated a new generation of POC glucose device, the HemoCue(®) Glucose 201DMRT (201DMRT), for its suitability for (tight) glycemic control. MATERIALS AND METHODS: This study was performed according to the CLSI/STARD criteria. The 201DMRT was compared to the laboratory hexokinase glucose method (Siemens Dimension Vista(®)). The variation among different POC devices and cuvette lot numbers was examined. Additionally, the influence of the partial pressure of oxygen and hematocrit on glucose measurement was investigated. RESULTS: The 201DMRT showed a good agreement with the laboratory reference method. This was examined using Deming regression analysis, percentage Bland-Altman plot and a modified Clarke-error grid. The total analytical error at the clinically critical glucose concentrations of 5.6, 7.0 and 11.1 mmol/L (101, 126 and 200 mg/dL) was 6.4%, 4.3% and 3.0%, respectively. The total error among the different POC devices and among different cuvette lot numbers was <6.5%. Glucose measurements on the 201DMRT were not affected by changes in partial pressure of oxygen, whereas changes in hematocrit had influence on the results (3.4% for every 0.10 L/L change in hematocrit). CONCLUSIONS: The 201DMRT device can be used for glycemic control based on analytical results presented. However, the clinical applicability for tight glycemic control must be confirmed in a clinical study.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.015
metaresearch head score (Gemma)0.017
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.015
Threshold uncertainty score0.078

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0150.017
Meta-epidemiology (narrow)0.0010.000
Meta-epidemiology (broad)0.0010.001
Bibliometrics0.0010.001
Science and technology studies0.0000.001
Scholarly communication0.0010.001
Open science0.0010.001
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0010.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.036
GPT teacher head0.346
Teacher spread0.311 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations16
Published2012
Admission routes1
Has abstractyes

Explore more

Same venueClinical Chemistry and Laboratory Medicine (CCLM)Same topicHyperglycemia and glycemic control in critically ill and hospitalized patientsFrench-language works237,207