MétaCan
Menu
Back to cohort
Record W2139152632 · doi:10.1117/12.646864

A comparison of single use and reusable small core sized holmium:YAG laser fibers: Is there a difference in the risk of fiber failure?

2006· article· en· W2139152632 on OpenAlexaff
Bodo E. Knudsen, Randolph D. Glickman, Joel M.H. Teichman

Bibliographic record

VenueProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2006
Typearticle
Languageen
FieldEngineering
TopicSolid State Laser Technologies
Canadian institutionsUniversity of British Columbia
Fundersnot available
KeywordsCore (optical fiber)HolmiumMaterials scienceFiber laserFiberLaserOptical fiberOptoelectronicsOpticsComposite materialPhysics

Abstract

fetched live from OpenAlex

This study tests the hypothesis that repeat steam sterilization will result in an increase in the rate of holmium:YAG laser fiber failure during bench testing in a series of commercially available single use and reusable small core sized (200-272 μm) holmium:YAG laser fibers. Single use and reusable small core-sized holmium laser fibers were tested. Single use fibers included the Dornier Lightguide Super 200, Dornier DUR Laser Fiber Single Use, Optical Integrity Scopesafe 272, IQinc. LLF200TG-D and LLF273TG-D, Boston Scientific AccuFlex 200 and AccuFlex 273, and Lumenis Slimline EZ 200. Reusable fibers included the Dornier DUR Laser Fiber Reusable, Lumenis Slimline 200, Sharplan 200, Laser Peripherals RBLF-200, IQinc. LFT273NT, and Convergent Optiview SMH1020F. A Lumenis VersaPulse 100 watt and a PowerSuite 20 watt holmium:YAG lasers were used. Fibers were bent to 180 degrees at a diameter of 1.5 cm. The laser was fired at 1.5J, 10Hz for 30 minutes or until fiber fracture. Reusable fibers were sterilized for a total of twenty cycles using the manufacturers' specifications and retested after every five completed cycles. No fiber fractured with bending alone. Two of three Dornier Super 200 single use fibers fractured repeatedly within several laser pulses. The Laser Peripherals RBLF-200 fiber fractured during initial but not repeat testing. One of three reusable Dornier DUR Laser Fibers failed during testing after the twentieth steam sterilization cycle. Five of five Boston Scientific Accuflex 200 failed at the SMA connector and did not couple well with the Lumenis Powersuite laser. The test hypothesis was not supported.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.005
metaresearch head score (Gemma)0.010
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.005
Threshold uncertainty score0.027

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0050.010
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.001
Bibliometrics0.0010.000
Science and technology studies0.0000.001
Scholarly communication0.0000.001
Open science0.0010.000
Research integrity0.0010.000
Insufficient payload (model declined to judge)0.0010.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.023
GPT teacher head0.233
Teacher spread0.211 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2006
Admission routes1
Has abstractyes

Explore more

Same venueProceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIESame topicSolid State Laser TechnologiesFrench-language works237,207