MétaCan
Menu
Back to cohort
Record W2155917554 · doi:10.1177/152660280000700205

First-Generation Aortic Endografts: Analysis of Explanted Stentor Devices from the EUROSTAR Registry

2000· article· en· W2155917554 on OpenAlexaff
Robert Guidoin, Yves Marois, Yvan Douville, Martin W. King, Martín Castonguay, A. Traoré, Maxime Formichi, L.E. Staxrud, Lars Norgren, Patrice Bergeron, Jean‐Pierre Becquemin, José María Egaña, Peter L. Harris

Bibliographic record

VenueJournal of Endovascular Therapy · 2000
Typearticle
Languageen
FieldMedicine
TopicAortic aneurysm repair treatments
Canadian institutionsUniversité LavalUniversity of ManitobaHôpital Saint-François d'Assise
Fundersnot available
KeywordsMedicineSurgery

Abstract

fetched live from OpenAlex

PURPOSE: To examine the structure and healing characteristics of chronically implanted Stentor endografts that were explanted due to migration, endoleak, thrombosis, or aneurysm expansion. METHODS: The devices were harvested following reoperation (n = 5) or autopsy (n = 1) with implantation times ranging from 13 to 53 months. Structural modifications to the metal components were examined using radiography, endoscopy, and magnetic resonance imaging (MRI). Specimens taken from components of the modular stent-grafts were examined histologically and with scanning electron microscopy (SEM) to assess healing behavior. Physical and chemical stability of the nitinol wires and woven polyester graft material was evaluated using SEM and electron spectroscopy for chemical analysis. RESULTS: Although the endografts were retrieved for a variety of reasons, they exhibited similar healing and structural modifications. The woven polyester sleeve showed evidence of yarn shifting and distortion, yarn damage, and filament breakage leading to the formation of openings in the fabric. The luminal surface endografts showed incomplete healing characterized by a poorly organized, nonadherent thrombotic matrix of variable thickness. Radiographic and endoscopic observations indicated that structural failure of the grafts, particularly in the main aortic component, was related to severe compaction and dislocation of the metallic frame due to suture breaks. Corrosion marks were observed on some nitinol wires in all devices. Chemical analysis and ion bombardment of the nitinol wires revealed that the surface concentrations of titanium and nickel were not homogenous. The first layer was composed of carbon or organic elements, followed by a stratum of highly oxidized titanium with a low nickel concentration; the titanium-nickel alloy lay beneath these layers. CONCLUSIONS: Although the materials selected for construction of endovascular grafts appears judicious, the assembly of these biomaterials into various interrelated structures within the device requires further improvement.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.002
metaresearch head score (Gemma)0.002
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Observational · Consensus signal: Observational
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.002
Threshold uncertainty score0.009

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0020.002
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0020.001
Science and technology studies0.0000.000
Scholarly communication0.0010.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0010.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.029
GPT teacher head0.270
Teacher spread0.241 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designObservational
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations119
Published2000
Admission routes1
Has abstractyes

Explore more

Same venueJournal of Endovascular TherapySame topicAortic aneurysm repair treatmentsFrench-language works237,207