Quantitatively accurate activity measurements with a dedicated cardiac SPECT camera: Physical phantom experiments

2015· article· en· W2280185376 on OpenAlexafffund
Amir Pourmoghaddas, R. Glenn Wells

Bibliographic record

VenueMedical Physics · 2015
Typearticle
Languageen
FieldMedicine
TopicMedical Imaging Techniques and Applications
Canadian institutionsUniversity of OttawaCarleton University
FundersUniversity of OttawaNatural Sciences and Engineering Research Council of CanadaHeart and Stroke Foundation of Canada
KeywordsImaging phantomSingle-photon emission computed tomographyNuclear medicineMedical imagingCardiac imagingMedical physicsPhysicsComputer scienceMedicineRadiologyArtificial intelligence

Abstract

fetched live from OpenAlex

PURPOSE: Recently, there has been increased interest in dedicated cardiac single photon emission computed tomography (SPECT) scanners with pinhole collimation and improved detector technology due to their improved count sensitivity and resolution over traditional parallel-hole cameras. With traditional cameras, energy-based approaches are often used in the clinic for scatter compensation because they are fast and easily implemented. Some of the cardiac cameras use cadmium-zinc-telluride (CZT) detectors which can complicate the use of energy-based scatter correction (SC) due to the low-energy tail-an increased number of unscattered photons detected with reduced energy. Modified energy-based scatter correction methods can be implemented, but their level of accuracy is unclear. In this study, the authors validated by physical phantom experiments the quantitative accuracy and reproducibility of easily implemented correction techniques applied to (99m)Tc myocardial imaging with a CZT-detector-based gamma camera with multiple heads, each with a single-pinhole collimator. METHODS: Activity in the cardiac compartment of an Anthropomorphic Torso phantom (Data Spectrum Corporation) was measured through 15 (99m)Tc-SPECT acquisitions. The ratio of activity concentrations in organ compartments resembled a clinical (99m)Tc-sestamibi scan and was kept consistent across all experiments (1.2:1 heart to liver and 1.5:1 heart to lung). Two background activity levels were considered: no activity (cold) and an activity concentration 1/10th of the heart (hot). A plastic "lesion" was placed inside of the septal wall of the myocardial insert to simulate the presence of a region without tracer uptake and contrast in this lesion was calculated for all images. The true net activity in each compartment was measured with a dose calibrator (CRC-25R, Capintec, Inc.). A 10 min SPECT image was acquired using a dedicated cardiac camera with CZT detectors (Discovery NM530c, GE Healthcare), followed by a CT scan for attenuation correction (AC). For each experiment, separate images were created including reconstruction with no corrections (NC), with AC, with attenuation and dual-energy window (DEW) scatter correction (ACSC), with attenuation and partial volume correction (PVC) applied (ACPVC), and with attenuation, scatter, and PVC applied (ACSCPVC). The DEW SC method used was modified to account for the presence of the low-energy tail. RESULTS: T-tests showed that the mean error in absolute activity measurement was reduced significantly for AC and ACSC compared to NC for both (hot and cold) datasets (p < 0.001) and that ACSC, ACPVC, and ACSCPVC show significant reductions in mean differences compared to AC (p ≤ 0.001) without increasing the uncertainty (p > 0.4). The effect of SC and PVC was significant in reducing errors over AC in both datasets (p < 0.001 and p < 0.01, respectively), resulting in a mean error of 5% ± 4%. CONCLUSIONS: Quantitative measurements of cardiac (99m)Tc activity are achievable using attenuation and scatter corrections, with the authors' dedicated cardiac SPECT camera. Partial volume corrections offer improvements in measurement accuracy in AC images and ACSC images with elevated background activity; however, these improvements are not significant in ACSC images with low background activity.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.002
metaresearch head score (Gemma)0.004
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.002
Threshold uncertainty score0.010

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0020.004
Meta-epidemiology (narrow)0.0010.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.001
Science and technology studies0.0000.001
Scholarly communication0.0010.000
Open science0.0010.001
Research integrity0.0010.000
Insufficient payload (model declined to judge)0.0010.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.162
GPT teacher head0.410
Teacher spread0.248 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations8
Published2015
Admission routes2
Has abstractyes

Explore more

Same venueMedical PhysicsSame topicMedical Imaging Techniques and ApplicationsFrench-language works237,207