SU-F-T-573: Evaluation of EBT-XD Radiochromic Films for Verification of SBRT and SRS Treatment Delivery

2016· article· en· W2466298700 on OpenAlexaff
Jean‐François Aubry, Karim Zerouali

Bibliographic record

VenueMedical Physics · 2016
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsCentre Hospitalier de l’Université de Montréal
Fundersnot available
KeywordsImaging phantomNuclear medicineIonization chamberRadiosurgeryDosimetryMaterials scienceDose profileCalibrationScannerIrradiationRadiation therapyMedicinePhysicsOpticsRadiology

Abstract

fetched live from OpenAlex

Purpose: To evaluate the accuracy and precision of radiochromic films EBT-XD for quality control of stereotaxic body radiation therapy (SBRT) and stereotactic radiosurgery (SRS) plan delivery. Methods: A film cut in 3×10 cm2 strips was irradiated from 0 to 20 Gy in increments of 1 to 1.5 Gy using a 15 MeV electron beam. Triple-channel film calibration was completed 24 hours later by scanning the film strips on an Epson 10000XL scanner using a well-defined protocol. Several dose measurements of increasing complexity were subsequently performed with Varian iX accelerators. Pieces of films were first irradiated in a solid water phantom with 6 MV photon beams and a static gantry to doses spanning the calibration range, either in a single field or multiple fields setup. High dose (>15 Gy per fraction) IMRT plans were then measured. Finally films were irradiated with volumetric modulated arc therapy (VMAT) plans of lung and spinal lesions with prescribed doses per fraction between 8 and 20 Gy. The dose measured with the films was compared to the calculated dose from the Eclipse planning system using the Anisotropic Analytical Algorithm (AAA). Results: 77 dose measurements were compared to either ion chamber measurements or dose calculations (reference). The average dose difference between film measurements and reference was 0.7 % and the standard deviation was 1.3%. The maximum and minimum dose differences were +3.5% and −2% in the 4 Gy to 20 Gy range. Measured dose profiles of lung and vertebra treatment plans agreed very well with the calculations. Conclusion: EBT-XD films are a useful dosimeter for quality control of SBRT and SRS plan delivery. The measurement of a full 2D dose plane with high spatial resolution and acceptable dose accuracy make it an advantageous choice compared to other detectors such as ion chambers or diodes.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.003
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.003
Threshold uncertainty score0.008

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.003
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0010.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0010.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.024
GPT teacher head0.320
Teacher spread0.296 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2016
Admission routes1
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced Radiotherapy Techniques→French-language works237,207→