SU‐F‐J‐52: A Novel Approach to X‐Ray Tube Quality Assurance for CBCT Systems in Order to Better Assess the Patient Imaging Dose in a Large, Multi‐Unit Treatment Facility

2016· article· en· W2468048118 on OpenAlexaff
L Buckley, Robert M. Webb, Carole Lambert, Balázs Nyíri, L Gerig

Bibliographic record

VenueMedical Physics · 2016
Typearticle
Languageen
FieldPhysics and Astronomy
TopicAdvanced Radiotherapy Techniques
Canadian institutionsOttawa Hospital
Fundersnot available
KeywordsQuality assuranceCalibrationImaging phantomMedical physicsNuclear medicineDetectorMedical imagingDosimetryImage qualityGenerator (circuit theory)Cone beam computed tomographyMonitor unitMedicineComputer scienceMathematicsPhysicsOpticsComputed tomographyRadiologyArtificial intelligenceStatistics

Abstract

fetched live from OpenAlex

Purpose: To standardize the tube calibration for Elekta XVI cone beam CT (CBCT) systems in order to provide a meaningful estimate of the daily imaging dose and reduce the variation between units in a large centre with multiple treatment units. Methods: Initial measurements of the output from the CBCT systems were made using a Farmer chamber and standard CTDI phantom. The correlation between the measured CTDI and the tube current was confirmed using an Unfors Xi detector which was then used to perform a tube current calibration on each unit. Results: Initial measurements showed measured tube current variations of up to 25% between units for scans with the same image settings. In order to reasonably estimate the imaging dose, a systematic approach to x‐ray generator calibration was adopted to ensure that the imaging dose was consistent across all units at the centre and was adopted as part of the routine quality assurance program. Subsequent measurements show that the variation in measured dose across nine units is on the order of 5%. Conclusion: Increasingly, patients receiving radiation therapy have extended life expectancies and therefore the cumulative dose from daily imaging should not be ignored. In theory, an estimate of imaging dose can be made from the imaging parameters. However, measurements have shown that there are large differences in the x‐ray generator calibration as installed at the clinic. Current protocols recommend routine checks of dose to ensure constancy. The present study suggests that in addition to constancy checks on a single machine, a tube current calibration should be performed on every unit to ensure agreement across multiple machines. This is crucial at a large centre with multiple units in order to provide physicians with a meaningful estimate of the daily imaging dose.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.002
metaresearch head score (Gemma)0.003
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: Bench or experimental
GenreCandidate signal: Empirical · Consensus signal: none
Teacher disagreement score0.002
Threshold uncertainty score0.011

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0020.003
Meta-epidemiology (narrow)0.0010.001
Meta-epidemiology (broad)0.0010.000
Bibliometrics0.0010.001
Science and technology studies0.0000.001
Scholarly communication0.0020.001
Open science0.0010.001
Research integrity0.0010.001
Insufficient payload (model declined to judge)0.0010.000

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.041
GPT teacher head0.342
Teacher spread0.301 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2016
Admission routes1
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced Radiotherapy Techniques→French-language works237,207→