Abstract

fetched live from OpenAlex

Purpose: The HexaPOD™ six degree of freedom couchtop is equipped with an optical tracking system, consisting of a stereoscopic camera and a reference frame (RF) carrying infrared reflective markers. The manufacturer recommends placing the RF within 50 cm from linac isocenter (ISO), which is a serious limitation since the RF does not fit around the shoulders of most brain patients. This study quantifies the impact of extended RF distances from ISO on positional accuracy. Methods: An in‐house tool with an estimated resolution of 0.3 mm and 0.1° was used. It is a large cube and a mathematical model of HexaPOD motion to determine the intersection of room lasers with the ruled cube edges. Combinations of translations (±1 and ±3 cm) and rotations (±2.5°) were executed on two HexaPOD couchtops for multiple RF distances from ISO (35 to 77 cm). For each combination, ten laser readings were fed into a least squares algorithm to determine the executed translations and rotations while minimizing operator reading errors. Results: The usable tracking volume is up to an RF distance of 82 cm from ISO. Positional accuracy of the HexaPOD/iGuide system is 0.6 mm and 0.1° (95% confidence). Positional accuracy variations versus RF distance from ISO are statistically insignificant (p = 0.05). Our results generally confirm recent internal estimates by the manufacturer (for future release). Conclusions: RF distances up to 77 cm from ISO are clinically acceptable, provided performing a patient safety study with a verification scan.

Fetched live from OpenAlex and de-inverted. Abstracts are not stored in this database: the inverted indexes are 8.6 GB of the frame’s 9.3 GB of text, and the host has 13 GB free.

How this classification was reachedexpand

Full frame machine prediction

Teacher imitation

Not calibrated prevalence, not ground truth. Human validation pending. The Gemma side is a direct model label for every work in the frame, read from the title-only record. The Codex side is a classifier learned from the 10,348 direct Codex labels and calibrated to design-weighted sample rates; fields without enough sample support carry no Codex call. Candidate is the union of the two sides; consensus is their intersection. These outputs are machine_predicted_unvalidated and are not human labels.

metaresearch head score (Codex)0.001
metaresearch head score (Gemma)0.006
Version: metacan-v3-hybrid-931329e0061cValidation status: machine_predicted_unvalidated
Candidate categoriesnone
Consensus categoriesnone
DomainCandidate signal: none · Consensus signal: none
Study designCandidate signal: Bench or experimental · Consensus signal: none
GenreCandidate signal: Empirical · Consensus signal: Empirical
Teacher disagreement score0.004
Threshold uncertainty score0.012

Distilled classifier scores by category (both heads)

CategoryCodexGemma
Metaresearch0.0010.006
Meta-epidemiology (narrow)0.0000.000
Meta-epidemiology (broad)0.0000.000
Bibliometrics0.0000.000
Science and technology studies0.0000.000
Scholarly communication0.0000.000
Open science0.0000.000
Research integrity0.0000.000
Insufficient payload (model declined to judge)0.0040.001

Machine scores (provisional)

The two teacher heads of the student model, read on this work. A score orders the frame for review; it never asserts a category, and the validation status ships verbatim with every row.

Baseline scores from an immature model (maturity gate not passed, 7 training rounds). Scores rank; they never assert a category.

Opus teacher head0.010
GPT teacher head0.310
Teacher spread0.300 · how far apart the two teachers sit on this one work
Validation statusscore_only:v0-immature-baseline · verbatim from the scoring run: score_only means the number may rank works, and no category label ships from it

Classification

machine, unvalidated

Machine predicted; a candidate call from one source (direct Gemma or distilled Codex), not a consensus.

The models applied no category: nothing in the taxonomy fit this work.
Study designBench or experimental
Domainnot available
GenreEmpirical

How this classification was reached, model by model and score by score, is at the end of the page under "How this classification was reached".

Quick stats

Citations0
Published2016
Admission routes1
Has abstractyes

Explore more

Same venueMedical Physics→Same topicAdvanced Radiotherapy Techniques→French-language works237,207→